Clean the compressor housing as follows:
• Wipe the outer surface of the
housing with a clean, damp
cloth.
4.2 Multiple patient use
Cleaning and disinfection cycles
Nebuliser (with
connection tubing) and
mask
Replace before each patient change
Compressor housing
Disinfect before each patient change
Air filter
Replace after 200 operating hours
(approx. 1 year)
Nebuliser and mask
The nebuliser and mask are not suitable for use by multiple
patients. Use a separate nebuliser (including connection
tubing) and a separate mask for each patient.
Disinfecting the compressor housing
For disinfection, use a standard, alcohol-based disinfectant
(e.g., isopropanol). Regarding the use and quantity of the disin-
fectant, it is imperative to follow the information for use of the
agent used.
To disinfect the compressor, proceed as follows:
– 30 –
Содержание COMPACT2
Страница 2: ......
Страница 40: ...40...
Страница 45: ...Limity o et ov n max 50 cykl o et ov n 45...
Страница 46: ...46...
Страница 47: ...Inhala n syst m PARI COMPACT2 Z RU N LIST Na kompresor poskytujeme 2 roky z ruky Z ru n doba za n datem prodeje...
Страница 50: ......
Страница 98: ...50...
Страница 102: ......
Страница 145: ...Processing limits max 50 processing cycles 45...
Страница 146: ...46...
Страница 150: ......
Страница 196: ...48...
Страница 197: ...49...
Страница 198: ...50...
Страница 202: ......
Страница 248: ...48...
Страница 249: ...49...
Страница 250: ...50...
Страница 254: ......
Страница 296: ...44...
Страница 302: ...50...
Страница 306: ......
Страница 354: ...50...
Страница 358: ......
Страница 406: ...50...