Housing dimensions
(W × H × D)
18.5 cm × 13.0 cm × 15.0 cm
Weight
1.55 kg
Pressure
5
1.2 bar
Compressor flow
5
4.1 l/min
Sound pressure level
56 dB(A)
Classification according to IEC 60601-1 /
EN 60601-1
Type of electric shock protection
Protection class II
Degree of protection from electric shock of
the applied part (nebuliser)
Type BF
Degree of protection according to IEC 60529 /
EN 60529 against penetration by water and
solid substances
IP 21
Degree of protection when used in the
presence of flammable mixtures of
anaesthetics with air, with oxygen, or with
nitrous oxide
No protection
Operating mode
Continuous
operation
Electromagnetic compatibility
Electrical medical equipment is subject to special precautionary
measures with regard to electromagnetic compatibility (EMC).
Such equipment must only be installed and operated in accord-
ance with EMC instructions.
Portable and mobile high-frequency communication devices
can disrupt electrical medical equipment. Using accessories,
converters and power cords other than those specified (with
the exception of converters and power cords that the
5) Towards nebuliser nozzle (Ø 0.48 mm).
– 34 –
Содержание COMPACT2
Страница 2: ......
Страница 40: ...40...
Страница 45: ...Limity o et ov n max 50 cykl o et ov n 45...
Страница 46: ...46...
Страница 47: ...Inhala n syst m PARI COMPACT2 Z RU N LIST Na kompresor poskytujeme 2 roky z ruky Z ru n doba za n datem prodeje...
Страница 50: ......
Страница 98: ...50...
Страница 102: ......
Страница 145: ...Processing limits max 50 processing cycles 45...
Страница 146: ...46...
Страница 150: ......
Страница 196: ...48...
Страница 197: ...49...
Страница 198: ...50...
Страница 202: ......
Страница 248: ...48...
Страница 249: ...49...
Страница 250: ...50...
Страница 254: ......
Страница 296: ...44...
Страница 302: ...50...
Страница 306: ......
Страница 354: ...50...
Страница 358: ......
Страница 406: ...50...