- 11 -

8.Set example of parameter

  8-1.Flow of parameter set 

            Parameter should set according to the flow as shown hereunder. 

In case setting up value over input range exceeding specification, setup value  
of parameter will indicate 「Err」on indication L E D, let you know setup value  
is wrong and do not accept next operation. 

In this case, please change to correct set value again. 

      8-2.Example of parameter set 
          Following is a set example to set integral time at 15 second. 
          (Providing that set value was 0 second before change.) 
          (1)  Push MOD button 1 time・・・・・・・・・・move to parameter window and LED(orange). 

written Range on the left will light on. 

          (2)  Push SEL button 5 times・・・・・・・・ LED(orange)lit goes down in turn 
                                                           and INTEGRATION TIME LED will light. 
          (3)  By using both 

 button and ↑button, 

               set Indication LED at「15」. 

          (4)  Push MOD button1 time・・・・・・・・・return to measure window and start measure 

again at integral15 second. 

          You operate other parameters with same procedure, please. 
          In case setting other parameters continuously, at (4) above instead of MOD you push 

SEL and can perform both setting parameter and moving to plural items at the same time. 

          Reference: Parameter and light up of corresponding LED 

RANGE             :  only RANGE LED light up 
L1 ON            :  only L1 ON LED light up 

                  L1 OFF           :  only L1 OFF LED light up 
                  L2 ON            :  only L2 ON LED light up 
                  L2 OFF           :  only L2 OFF LED light up 
                  INTEGRATION TIME :  only INTEGRATION TIME LED light up 

9.Maintenance  

 

    Measure probe should regularly be wiped and cleaned with dry rag. 

You please keep it cleaned once in every 3 months but more often because of changing media 
or measure environment, you see. 
When cleaning, you please switch off power. 
Temperature of pipe may be high and dust monitor might be heated accordingly. 
Be careful! Or you might get your hands burned.  

Caution: The earth line should not be removed even at the maintenance in case 

               it may cause a fault. 

Measure 

window 

Parameter mode 

Value set 

Parameter select 

MOD button 
Press down 
(to    parameter 
mode) 

SEL button 
Push down 
Respective 
parameter select 

Move  beam  by 

button 
↑ Each  button 
push down change 
value 

Push down MOD button  
(after parameter set, to  measure 
window) 

SEL button push down 
(after set parameter move 
to other parameter item) 

!

Содержание PFM-KCU11

Страница 1: ...e 11 10 Trouble shooting 12 13 INSTRUCTION MANUAL FOR Dust Monitor TYPE PFM KCU11 Transducer PFM M11P Sensor PFM M11PT High temperature type 1 8 18 Norimatsu Higashi Yahatanishi ku Kitakyushu 807 0837 Japan Phone No 8193 691 3731 Fax No 8193 691 3735 http www matsushima m tech com E mail sales matsushima m tech com The operator should read this Instruction Manual carefully and handle the device co...

Страница 2: ...y long wire Tighten the cover lead outlet etc properly so that dust rainwater etc do not enter inside the equipment Do not use the equipment under a corrosive condition NH3 SO2 CI2 etc Be sure to tighten the cable grand so that outer air does not enter inside the equipment When applying piping connection such as conduit etc instead of cable grand apply putty or equivalents on the cable entry so th...

Страница 3: ...ry under that environment As to considering density regarding to Dust Monitor of Friction Electric Charge Mechanism it requires you to consider influence due to variation of particle characteristics as afore mentioned and it does not always meet density specified Density indicated on specification is for design development of machinery and is publicized for you customers to use as a guideline in d...

Страница 4: ...rature 20 65 Transducer 50 1 2 8 2 Humidity 95 RH Condition State of case sealing up 2 9 Keeping condition 2 9 1 Temperature 20 65 2 9 2 Humidity 95 RH 2 10 Output signal 2 10 1 Current output DC 4 20mA isolated output load resistance 500Ω max 23 3mA 2 10 2 Contact output Alarm contact output 1 250VAC 2A 30VDC 2A Upper limit contact limit output 2 250VAC 2A 30VDC 2A 2 11 Fitting method R1 fit scre...

Страница 5: ...US316L 3 Radiation fin Alminium φ98 30 85 Lead outlet G1 2 6 Transducer PFM KCU11 Sensor PFM M11PT 1 Please detach A between socket and insulator 10mm or more 2 Refer to the specifications for length L1 L2 3 The probe should be touched directly by hand without grounding while being energized 4 Please be sure to connect earth terminal 5 Please tighten the Bolt and Nut when you install it 6 Please m...

Страница 6: ...rotection and is not a raincoat Attention Exclusive cable is transmitting very weak signal Cable should be protected with by all means steel made electric wire pipe independently so that inductive noise form ambient area does not bother signal Securely close the lead outlet to prevent outside air from entering in the main unit Do not remove the accessory Silica gel for dehumidification Where using...

Страница 7: ... for Transducer attaching when you remove the attaching screw covers Please loosen these joining screws of the terminal block part with a Phillips driver You can see the terminal block of the Transducer when you remove the main block part from the terminal block part after having loosened these Joining Screws Joining Screws Mtg Hole Mtg Hole Mtg Hole Mtg Hole Fig 7 Joining screws of terminal and m...

Страница 8: ...2 Completion drawing 1 85 3 5 4 Packing 1 Please pass Cap Clamping member and Rubber bush through the cable Please peel off the cable covering from the verge by the length of 85mm 2 Please make the shield line asunder Please wrap an asunder shield line as shown in left figure 3 Please pass Shield ring through the cable Please peel off the cover of the core wire by 5mm Please use solderless termina...

Страница 9: ...ER LED Light on as power on FAULT LED CPU on when abnormal L1 ON LED Light on over limit1set L2 ON LED Light on over limit1set 4 beams indicate LED detection density will indicate per 0 100 complying with setup for button use when changing beam of number value use when increasing number value MOD use when switching parameter set mode and measure mode Pushing after set change will memorize paramete...

Страница 10: ...output due to its output linear Reference Talc Talc is non organic powder made from ore called Talc being crushed into gray white particle and smooth powder Talc with its chemical name hydrous magnesium silicate Mg3Si4O10 OH containing main ingredient of SiO2 approx 60 MgO approx 30 and combined water of 4 88 As its physical nature Talc of true specific gravity 2 7 2 8 is a substance having lowest...

Страница 11: ...Customers expertise respectively Limit 1 2 and action chart of trouble alarm Limit output upper limit alarm Power OFF to open COM NO Trouble alarm Power OFF to close COM NO 20mA 4mA Time Range to set cang Limit set value ON Limit set value OFF Hysteresis Trouble Device normal Device normal COM NO close LED On COM NO open LED Off COM NO open LED Off COM NO close LED on COM NO open LED off COM NO op...

Страница 12: ... above instead of MOD you push SEL and can perform both setting parameter and moving to plural items at the same time Reference Parameter and light up of corresponding LED RANGE only RANGE LED light up L1 ON only L1 ON LED light up L1 OFF only L1 OFF LED light up L2 ON only L2 ON LED light up L2 OFF only L2 OFF LED light up INTEGRATION TIME only INTEGRATION TIME LED light up 9 Maintenance Measure ...

Страница 13: ...ng is wrong rewire correctly Remove Sensor and check if any sticking on Probe electrode If yes power off and wipe them off with moist cloth See NOTE Remove Sensor and check if Probe electrode has dew condensation If yes power off and wipe it off with moist cloth See NOTE Check if an applicable amount is flowing If necessary change the measurement range and integral setting Output indication goes o...

Страница 14: ...ength and check if they are not in contact If yes change electrode length See NOTE Consult with Sales Dept to check the soundness of Device Power OFF then ON Fault LED is lighted Wiring is wrong The supplied power supply is out of allowable range CPU abnormal Check wiring of power output signal lines exclusive cable See Page 6 5 How to connect If wiring is wrong rewire correctly Check power supply...

Отзывы: