D.11 Simulation mode
Simulation mode
allows the user to test the FF
implementation without the instrument providing real data.
The meter PCB is shipped with simulation mode disabled.
To enable simulation mode, complete the following steps:
CAUTION!
To prevent damage to the electronic
components, always use
ESD
protection
whenever handling printed circuit boards.
1. Remove the PCB from the meter.
2. Locate the jumper
P5
(see
figure 34
below).
3. Move the
P5
jumper to the left one place (pins 2 and 3)
to enable simulation mode.
4. Re-install the PCB into meter
5. Verify that the
block
error field in the
resource block
shows
SimulationActive
status.
Jumper P5
Figure 34: P5 on XMT1000 PCB
65