1.5.4 Wiring the CH1 and CH2 Flying Lead
Refer to Figure 6 on page 6 and Figure 7 on page 7 to locate
the Channel1 and Channel 2 flying lead terminal blocks.
To wire the CH 1 flying lead to the terminal block 5, refer to
Figure 8 on page 9 and complete the following steps:
1. Disconnect the main power to the unit and remove the
rear cover.
2. Install the required cable gland in the chosen conduit
hole on the side of the electronics enclosure.
3. Feed one end of the CH 1 downstream RG62 cable red
wire through the conduit hole of DN terminal, and wire it
to the terminal block.
4. Feed one end of the CH 1 downstream RG62 cable black
wire through the conduit hole of RTN terminal below the
DN terminal, and wire it to the terminal block.
5. Feed one end of the CH 1 upstream RG62 cable red wire
through the conduit hole of UP terminal, and wire it to the
terminal block.
6. Feed one end of the CH 1 upstream RG62 cable black wire
through the conduit hole of RTN terminal above the UP
terminal, and wire it to the terminal block.
7. If wiring of the unit has been completed, reinstall the
wiring cover on the enclosure and tighten the set screw.
To wire the CH 2 flying lead to the terminal block 6, repeat
the above steps.
1.5.5 Wiring the Line Power
The XMT1000 may be ordered for operation with power
inputs of either 100-240 VAC or 12-28 VDC. The label on the
side of the enclosure lists the meter’s required line voltage
and power rating. Be sure to connect the meter to the
specified line voltage only.
Figure 8: CH1 and CH2 Flying Lead Wiring
9