A.2 Electronics

Enclosure

Standard:

 Epoxy-coated aluminum weatherproof 4X/IP66 

Class I, div 1, groups B, C and D, flameproof II 2 G Ex d IIC  

T5/T6

Optional:

 Stainless steel

Dimensions (standard)

Weight:

 10 lb (4.5 kg)

Size 

(h x d): 8.2 in. x 6.6 in. (208 mm x 168 mm)

Paths

1, 2 or 3 paths

Display

128 x 64 mono-color LCD display, configurable for single  

or dual measurement parameters

Display languages

English

Keypad

Built-in magnetic, six-button keypad, for full  

functionality operation

Inputs/outputs

Standard: 

one analog output**, one digital output*,  

service/modbus (RS485)

Optional: 

one analog output** with HART

*Digital outputs are programmable as either pulse or 

frequency outputs
**Analog outputs are NAMUR NE43 compliant

Power supplies

Standard:

 100-240 VAC (50/60 Hz)

Optional:

 12 to 28 VDC

Power consumption

15 watts maximum

Wiring connection

Conduit entries include 6 x 3/4” NPT and 1 x 1/2” NPT on 

bottom, consult Panametrics for available adapters

Electronics classifications (pending)

USA/Canada

 - explosion-proof class I, division 1,  

groups B, C, and D 

ATEX

 - Flameproof II 2 G Ex d IIC T6 Gb

IECEx

 - Flameproof Ex d IIC T6 Gb 

RoHS

 

compliance

 (Directive 2011/65/EU)

CE marking

 (EMC directive 2014/30/EU, LVD 2014/35/EU) 

WEEE compliance

 (Directive 2012/19/EU)

Electronics mounting

Standard: 

Local mounting (on meter body)

Optional:

 Remote mounting (recommended for  

process temperatures exceeding 150°C). The maximum 

distance is 1000 ft (300 m) using RG-62 coaxial cable  

or equivalent. If longer distances are required, consult  

the factory for assistance.

Terminal blocks

Table 5: Standard terminal block (output A)

I/O type

Connection

Specifications

Analog  

output

Active output

Output current: 0-22 mA
Max load: 600 Ω

Pulse, 

frequency

Active output

Output voltage:  

5 VDC 
Max voltage with  

light load: 7 VDC

RS485 

modbus

RS485 

communications

Standard RS485 

communications

Operating temperature

–40° to 140°F (–40° to +60°C)

Note: 

The LCD display is only visible down to -13°F (-25°C).

Storage temperature

–40° to 158°F (–40° to 70°C)

Humidity (operating and storage)

10-90% RH

25

Содержание PanaFlow XMT1000

Страница 1: ...PanaFlow XMT1000 User s manual 910 313 Rev A...

Страница 2: ...ii...

Страница 3: ...PanaFlow XMT1000 Panametrics liquid flow ultrasonic transmitter User s manual 910 313 Rev A June 2017 iii...

Страница 4: ...no content intended for this page iv...

Страница 5: ...mounting 4 1 4 6 Cable lengths 4 1 4 7 Transducer cables 4 1 5 Making the electrical connections 5 1 5 1 Wiring the analog outputs 5 1 5 2 Wiring the digital output 6 1 5 3 Wiring the modbus service p...

Страница 6: ...D 2 1 Network configuration 51 D 2 2 Polarity 51 D 2 3 Connection 51 D 2 4 FISCO fieldbus intrinsically safe concept 52 D 2 5 DD file 52 D 2 6 Default node address 52 D 3 Specifications 53 D 3 1 Gene...

Страница 7: ...e following link https www bakerhughesds com sales terms conditions Typographical conventions Note These paragraphs provide information that provides a deeper understanding of the situation but is not...

Страница 8: ...tion Make sure that unauthorized personnel cannot gain access to the operation of the equipment Environmental compliance RoHS The PanaFlow XMT1000 fully complies with RoHS regulations Directive 2011 6...

Страница 9: ...European customers To meet CE marking requirements all cables must be installed as described in CE marking compliance 1 2 CE marking compliance For CE marking compliance or installation in high noise...

Страница 10: ...r Care immediately for assistance The XMT1000 is supplied with both a serial number label and a certification label for identification of the instrument see figure 1 below and figure 2 The system can...

Страница 11: ...XMT1000 flow transmitter in a location isolated from vibrations Avoid installing it near equipment that generates low frequency high energy random vibrations 1 4 3 Sunlight exposure The installer shou...

Страница 12: ...oof enclosure Typically the enclosure is mounted as close as possible to the transducers When choosing a site for a remote mount installation which is recommended for process temperatures exceeding 15...

Страница 13: ...unterclockwise until it comes free from the enclosure 4 Install any required cable glands in the appropriate conduit holes on the opposite side of the enclosure 5 Note the labels inside the rear cover...

Страница 14: ...Pin Description 1 RS485 modbus 2 RS485 modbus 3 RS485 common 4 Frequency totalizer cal 5 Frequency totalizer cal 6 Analog 4 20MA OUT 7 Analog 4 20MA RTN Additional customer ground 3x PCB ground Digit...

Страница 15: ...tightened before applying power in a hazardous environment Note Prior to use the analog output must be set up and calibrated Proceed to the next section to continue the initial wiring of the unit Note...

Страница 16: ...e cable through the conduit hole wire it to the terminal block 4 If wiring of the unit has been completed reinstall the wiring cover on the enclosure and tighten the set screw WARNING Make sure all co...

Страница 17: ...terminal below the DN terminal and wire it to the terminal block 5 Feed one end of the CH 1 upstream RG62 cable red wire through the conduit hole of UP terminal and wire it to the terminal block 6 Fe...

Страница 18: ...netic keypad are used to program the XMT1000 confirms the choice of a specific option and data entry within that option enables users to exit from a specific option without entering unconfirmed data a...

Страница 19: ...tion protocol developed by Modicon for use with its programmable logic controllers It is a method used for transmitting information over serial lines between electronic devices Pulse output In pulse t...

Страница 20: ...Menu maps Use the menu maps in this section to program the desired XMT1000 features Measurement display menu map rev 10 Main menu map rev 10 SYSTEM menu map rev 10 MAIN Board I O menu map rev 10 OPTI...

Страница 21: ...rd is lost meter must be reset Contact the factory for instructions Main menu Diagnostics Program Keypad lockout Totalizer 1 variable 2 variables Forward total Reverse total Net total Totalizer time D...

Страница 22: ...zer Density Actual volumetric Standard volumetric Flow units m s ft s Flow limits SoS low limit SoS high limit Density Measure settings Zero cutoff Averaging Actual volumetric Backlight Always on Time...

Страница 23: ...tual volumetric Standard volumetric Inventory total Analog output Error handling Full value Base value Percentage of scale Digital output Pulse Frequency Off Pulse value Error handling Measurement Tes...

Страница 24: ...d FF HART HART info Hardware revision Software revision Analog output Analog measurement Actual volumetric Standard volumetric Velocity Mass Percentage of scale Calibrate 4mA 20mA Base value Error han...

Страница 25: ...admin password 0 2 0 5 1M 2M 4M Pipe OD Pipe ID WT Channel configure Ch 3 Ch 1 Ch 2 In mm Pipe OD ID In WT mm WT In ID mm Pipe ID Wall thickness Path Length Axial Standard Special Frequency Static Tw...

Страница 26: ...UP CALIBRATION ADVANCED Operator admin password Calibration type Calibration points Calibration config Calibration run Calibration table Calibration factor Velocity Composite Off Gate Frequency Reynol...

Страница 27: ...og Reset settings Flash update On Off Hi Low View log Reset log Reset to commission Reset to factory Main board update Corr peak limit Velocity low limit Velocity high limit Acceleration limit Amp dis...

Страница 28: ...or information 3 1 1 Error header Table 2 Error header Error pattern Error header Communication error Cn n is error number Flow error En n is error number System error Sn n is error number XMIT error...

Страница 29: ...he same as E6 or E5 E5 Amplitude The signal amplitude exceeds the programmed limits This error may occur due to high signal attenuation or amplification due to changes in fluid properties or transduce...

Страница 30: ...trength is insufficient for reliable measurements The fluid sound speed must not vary excessively The XMT1000 will tolerate relatively large changes in the fluid sound speed as may be caused by variat...

Страница 31: ...e for roughness and or pitting Any transducer damaged in this manner must be replaced Contact Panametrics for information on transducers in materials suitable for the application Internal damage An ul...

Страница 32: ...Volumetric flow mass flow flow velocity and totalized flow Flow accuracy velocity Up to 0 3 of reading achievable when supplied with a complete flow meter system and process calibration Accuracy depen...

Страница 33: ...ics for available adapters Electronics classifications pending USA Canada explosion proof class I division 1 groups B C and D ATEX Flameproof II 2 G Ex d IIC T6 Gb IECEx Flameproof Ex d IIC T6 Gb RoHS...

Страница 34: ...ime objective of the flow meter is to measure flow so the modbus server has a low priority B 2 Modbus register map Table 6 XMT1000 modbus register map revision 4 19 Reg in hex Access level Register ID...

Страница 35: ...Unit_StdVol Global unit group 14 for standard volumetric Unitless RW 4 INT32 51C Operator eUnit_Therm Global unit group 15 for therm Unitless RW 4 INT32 51E Viewer eUnit_TotTime Global unit group 16 f...

Страница 36: ...r addr Unitless RW 4 INT32 588 Operator ePCModbus_Bits PC MODBUS bits per character Unitless RW 4 INT32 58A Operator ePCModbus_ Termination PC MODBUS termination Unitless RW 4 INT32 5C0 5C0 Operator e...

Страница 37: ...IdMin NetworkID_min Unitless RO 4 INT32 704 N A eSystem_User password General user password Unitless RO 4 INT32 706 N A eSystem_Admin password Admin user password Unitless RO 4 INT32 708 N A eSystem_M...

Страница 38: ...0 900 Operator eDisplay_Language Display language Unitless RW 4 INT32 902 Viewer eDisplay_Var1_value Display variable_ 1 register address Unitless RW 4 INT32 904 Viewer eDisplay_Var1_Unit Display vari...

Страница 39: ...Display_Select_ Decimal_1ST Select 1st display decimal Unitless RW 4 INT32 924 Viewer eDisplay_Type Type of DISPLAY Unitless RW 4 INT32 926 Operator eDisplay_Timeout TimeOut for DISPLAY 13 RW 4 INT32...

Страница 40: ...ut_max Maximum timeout for DISPLAY 13 RO 4 INT32 B02 N A eDisplay_Timeout_Min Minimum timeout for DISPLAY 13 RO 4 INT32 B04 N A eDisplay_contrast_max Maximum contrast for DISPLAY Unitless RO 4 INT32 B...

Страница 41: ...lue percent of span 12 RW 4 IEEE 32 bit 1008 Operator eAout1_BaseValue Analog out 1 base value 1 9 14 17 20 RW 4 IEEE 32 bit 100A Operator eAout1_FullValue Analog out 1 full value 1 9 14 17 20 RW 4 IE...

Страница 42: ...W 1030 1030 Operator eAout4_ErrValue Analog out 4 error handling value 8 RW 4 IEEE 32 bit 1032 Operator eAout4_ZeroValue Analog out 4 zero 8 RW 4 IEEE 32 bit 1034 Operator eAout4_spanValue Analog out...

Страница 43: ...tor eAout3_Mode Analog out 3 operating mode Unitless RW 4 INT32 11A2 Operator eAout3_Type Analog out 3 type Unitless RW 4 INT32 11A4 Operator eAout3_Adress Analog out 3 measurement register address Un...

Страница 44: ...N A eAout1_spanValue_ max Maximum analog out 1 span 8 RO 4 IEEE 32 bit 1606 N A eAout1_TestValue_ max Maximum analog out 1 test value percent of span 12 RO 4 IEEE 32 bit 1608 N A eAout1_BaseValue_ max...

Страница 45: ...an 8 RO 4 IEEE 32 bit 1626 N A eAout3_TestValue_ max Maximum analog out 3 test value percent of span 12 RO 4 IEEE 32 bit 1628 N A eAout3_BaseValue_ max Maximum analog out 3 base value 1 9 14 17 20 RO...

Страница 46: ...m analog out 1 test value percent of span 12 RO 4 IEEE 32 bit 1A08 N A eAout1_BaseValue_Min Minimum analog out 1 base value 1 9 14 17 20 RO 4 IEEE 32 bit 1A0A N A eAout1_FullValue_Min Minimum analog o...

Страница 47: ...IEEE 32 bit 1A28 N A eAout3_BaseValue_Min Minimum analog out 3 base value 1 9 14 17 20 RO 4 IEEE 32 bit 1A2A N A eAout3_FullValue_Min Minimum analog out 3 full value 1 9 14 17 20 RO 4 IEEE 32 bit IO...

Страница 48: ...ut_1 control output value 17 RW 4 IEEE 32 bit MAIN Dout1 int RW 2100 2100 Operator eDout1_PulseTestValue Output_1 test pulse value Unitless RW 4 INT32 2102 Operator eDout1_freqFullfreq Output_1 freque...

Страница 49: ...nitless RW 4 INT32 2196 Operator eDout1_controlState Output_1 control output state Unitless RW 4 INT32 2198 Operator eDout1_controlType Output_1 control output type Unitless RW 4 INT32 219A Operator e...

Страница 50: ...IEEE 32 bit 2406 Operator eDout2_freqFullValue Output_2 frequency full value 1 9 14 17 20 RW 4 IEEE 32 bit 2408 Operator eDout2_AlarmValue Output_2 alarm value 1 9 14 17 20 RW 4 IEEE 32 bit 240A Oper...

Страница 51: ...rator eDout2_AlarmState Output_2 alarm state Unitless RW 4 INT32 258E Operator eDout2_AlarmType Output_2 alarm type Unitless RW 4 INT32 2590 Operator eDout2_AlarmAddress Output_2 alarm measurement reg...

Страница 52: ...out2 real RO 2600 TBD Dout2 int RO 2700 Dout max real RW 2800 Dout max int RW 2900 MAIN Dout max real RO 2A00 2A00 N A eDout1_PulseValue_max Maximum output_1 pulse value 17 RO 4 IEEE 32 bit 2A02 N A e...

Страница 53: ..._ max Maximum output_2 control output value 17 RO 4 IEEE 32 bit MAIN Dout max int RO 2B00 2B00 N A eDout1_Pulse TestValue_ max Maximum output_1 test pulse value Unitless RO 4 INT32 2B02 N A eDout1_fre...

Страница 54: ...bit 2E08 N A eDout1_Alarm Value_Min Minimum output_1 alarm value 1 9 14 17 20 RO 4 IEEE 32 bit 2E0A N A eDout1_controlValue_Min Minimum output_1 control output value 17 RO 4 IEEE 32 bit TBD Dout Min r...

Страница 55: ...06 N A eDout1_freq ErrValue_Min Minimum output_1 frequency error handling value 6 RO 4 INT32 TBD Dout Min int RO 2F80 2F80 N A eDout2_Pulse TestValue_ Min Minimum output_2 test pulse value Unitless RO...

Страница 56: ...N A eFFDensity_reading Measurement for fieldbus for density Fieldbus unit type register for density RO 4 IEEE 32 bit 3602 N A eFFVolumetric_reading Measurement for fieldbus for volumetric Fieldbus un...

Страница 57: ...be used to disable write access to the instrument via HART This slide switch pictured in figure 17 below is designed to lock out HART configuration access for those customers who require this extra le...

Страница 58: ...ype TV value QV value PV loop current Login Service menu Logout Review menu Squawk Clear totals Use preference menu I O options menu Sensor setup menu Meter review HART review HART output menu Main an...

Страница 59: ...criptor Universal rev Tag HART settings Dev ID Software revision of required preambles Hardware revision Field device revision Date Model TV value TV measurement TV measurement class Manufacturer TV r...

Страница 60: ...l over the total value of the maximum current consumed by all devices including the host Conventional DC current cannot be used as is Terminators fieldbus requires two terminators Refer to the supplie...

Страница 61: ...t be installed if the fieldbus application is FISCO The FISCO cover should come installed from the factory as shown in figure 21 below The fieldbus cables should be installed through the XMT1000 upper...

Страница 62: ...2 VDC D 3 3 Communication Stack manufacturer Softing AG Backup LAS capable Yes Total number of VCRs 24 Fixed VCRs for configuration 1 LAS means Link Active Scheduler It can schedule a network if the m...

Страница 63: ...r block is in active mode To Enter configuration mode complete the following steps 1 Select the resource block others tab 2 Select SYSTEM_OPT_TYPE and set to option FI 3 Select SYSTEM_REQ_LEVEL and se...

Страница 64: ...this appendix or contact Panametrics customer service for assistance Check This category means the output of the device is invalid due to on going work on the device such as programming etc Maintenan...

Страница 65: ...river failure Failed Watch dog failure Failed DSP hardware errors Failed Default ISR DSP exception Failed DSP exception Failed Modbus communication loss No Modbus communication Failed Sensor value out...

Страница 66: ...contains parameters that can be transmitted onto the fieldbus via the AI block The user can view real time data and select the units for each of the parameters see figure 27 below Figure 27 Measureme...

Страница 67: ...ity g m3 kg L g ml kg m3 lb in3 lb ft3 lb gal Volumetric flow act m3 s m3 m m3 h m3 d L s L min L h ML d CFS CFM CFH ft3 d gal s GPM gal h gal d ImpGal s ImpGal min ImpGal h ImpGal d bbl s bbl min bbl...

Страница 68: ...mable parameters that are common to all three paths Figure 28 below shows the composite transducer block and table 9 lists the measurements and parameters that are available Note The R W designation m...

Страница 69: ...ESS R W CORR_PEAK_LOW_LIMIT R W ANALOGOUT_PERCENTSCALE R W Table 9 Available measurement values and parameters in the composite TB Composite TB measurements and parameters Measure ment Para meter ACCE...

Страница 70: ...tals can be controlled through foundation fieldbus see figure 29 below The user can start stop or reset batch totalizers by setting the option on the TOTALIZER_CMD function of the composite transducer...

Страница 71: ...rs Measure ment Parameter CH_SOUND_SPEED R CH_TRANSIT_TIME_UP R CH_TRANSIT_TIME_DN R CH_DELTA_T R CH_UP_SIGNAL_QUALITY R CH_DN_SIGNAL_QUALITY R CH_UP_AMP_DISC R CH_DN_AMP_DISC R Table 10 Available mea...

Страница 72: ...the error status on the fieldbus along with the real data The error status can be seen in the CH_x_reported error parameter on the channel transducer block In addition the quality parameter shown with...

Страница 73: ...UTION To prevent damage to the electronic components always use ESD protection whenever handling printed circuit boards 1 Remove the PCB from the meter 2 Locate the jumper P5 see figure 34 below 3 Mov...

Страница 74: ...ion of terminator requirements A value cannot be written to a parameter in the XMT1000 FF Not in configuration mode Enter correct admin password in resource block verify S1 configuration mode appears...

Страница 75: ...Genii models Image Model PN Description Key Benefits DPI620G FF Genii advanced modular calibrator and HART fieldbus communicator Fully featured communicators for device configuration trimming and cali...

Страница 76: ...rror display in the user interface v 19 Error handling vi 10 63 Error header 19 Error string communication v 19 Error string flow v 20 F FF revision vi 54 Fieldbus troubleshooting guide vi 65 Flow acc...

Страница 77: ...performance vi 23 Optional pc software 23 P Unpacking 1 2 Wiring diagram 48 Passcodes 9 Password vi 54 Paths 24 Personal safety equipment viii Physical damage 22 PID block vi 63 Pipe 21 23 Pipe probl...

Страница 78: ...aphical conventions vii U Unauthorized operation viii Units vi 25 57 58 Unpacking 1 2 Up arrow key 9 V Vibration exposure v 3 Vitality PC 23 Voltage inputs 7 9 W Warranty vii WEEE directive viii Wirin...

Страница 79: ...71...

Страница 80: ...uction targets across critical industries including Oil Gas Energy Healthcare Water and Wastewater Chemical Processing Food Beverage and many others Join the conversation and follow us on LinkedIn lin...

Отзывы: