5.0 Installing cable glands (cont.)
13. After the coaxial cables have been properly installed in their cable glands, they must be connected to the terminal blocks
inside the flow transmitter enclosure (see Figure 16 below) as follows:
Attention!
Because the PanaFlow LZ system has the flexibility to use several different Panametrics flow transmitters,
the photographs used in this manual may appear slightly different from your system. However, the basic
instructions are valid for all PanaFlow LZ systems.
a
. Connect the transducer wires to their terminal block per the instructions in the for your flow transmitter.
b.
Make sure all of the transducer wires are properly dressed and that they are neatly arranged inside the flow
transmitter enclosure.
c.
Use the flexible strain reliefs inside the flow transmitter enclosure to keep all of the transducer wires clear of the
threaded areas.
Figure 16: Terminating the transducer cables
11