DewPro® MMR30 Installation & Operation Manual
25
Chapter 4. Optional Display/User Interface
4.4.6
Directions for Detecting the Correct Error Code (cont.)
Table 19: Error Code Conversion Chart
0
0000
1
0001
2
0010
3
0011
4
0100
5
0101
6
0110
7
0111
8
1000
9
1001
10
1010
11
1011
12
1100
13
1101
14
1110
15
1111
Example 1:
Error Code: 1
1.
1/16 =
0.0625
No temperature error
2.
0.0625 x 16 =
1
Moisture error code = 1
3.
Error Code Bit Map
(0) (1)
0000 0001
+ >> RH < 0% (gas too dry)
Example 2:
Error Code: 39
1.
39/16 =
2.4375
Temperature error code = 2
2.
0.4375 x 16 =
7
Moisture error code = 7
3.
Error Code Bit Map
(2) (7)
0010 0111
++
>> Moisture Unit Selected
Depends on Temperature
+ >>
Output 1 < 4mA
+ >>
Temperature > + 85°C
Содержание DewPro MMR30
Страница 1: ...panametrics com Moisture A40251524 Rev B July 2020 DewPro MMR30 Installation Operation Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 8: ...Contents iv DewPro MMR30 Installation Operation Manual...
Страница 22: ...Chapter 3 Wiring Instructions 14 DewPro MMR30 Installation Operation Manual no content intended for this page...
Страница 36: ...Chapter 5 Troubleshooting 28 DewPro MMR30 Installation Operation Manual no content intended for this page...
Страница 40: ...Chapter 7 Accessories 32 DewPro MMR30 Installation Operation Manual no content intended for this page...
Страница 44: ...Warranty 36 DewPro MMR30 Installation Operation Manual no content intended for this page...