Chapter 2. Installation Guidelines
8
DewPro® MMR30 Installation & Operation Manual
2.4
Method III - No Flow Restriction
2.4.1
Low Pressure Closed Loop
Figure 7: DewPro with No Flow Restriction
Closed loop systems, with dew points >-15 °C, may operate at a low pressure. The air passing through the DewPro
bypass is fed back to the main stream after a pressure drop in the main line.
In this configuration, the DewPro bypass has no flow restriction at the inlet and outlet. The outlet is equipped with a
6 mm (1/4”) tube fitting to allow simple connection of the loop tubing.