Page
35
of
48
5.8 iCELLis 500+ Control System Electrical Installation
WARNING!
High voltage!
Always make sure this equipment is properly grounded.
WARNING!
Risk of injury to personnel and/or damage to equipment!
Only use the mains power supply cable that is supplied with the unit or that which carries equivalent
specification.
ALERT!
Risk of damage to equipment!
Never position the iCELLis 500+ control system so that it is difficult to access the machine’s mains/power
socket. A minimum clearance of 650 mm is required on each side of the iCELLis 500+ control system.
ALERT!
Risk of damage to equipment!
Never block the cooling fan outlets on the iCELLis 500+ control system when the iCELLis 500+ control
system is powered.
Electrical Power Requirements
Uninterruptible Power Supply (UPS) can be used with the iCELLis 500+ control system according to power
supply specifications mentioned in this IFU. The choice of UPS device is at the discretion of the user.
The iCELLis 500+ control system is supplied with its main power supply cable. Only use the cable
supplied with the unit. Insure that the iCELLis 500+ control system is connected to suitable overcurrent
protection device. Protective grounding for the system is provided through the ground terminal on the
plug provided.
Power requirements for iCELLis 500+ control system are displayed in Table 5.
Table 5: iCELLis 500+ control system power requirements
Input voltage (worldwide)
208*/230VAC – 50/60Hz
Maximum power consumption
4,000W
*On a 3 phase 120 Vac network, 208 Vac can be obtained by using the voltage between 2 phases.
Power Connection and Powering ON
PALL recommends turning OFF the iCELLis 500+ control system after each batch, or whenever the
system is not in use for a long period of time, to ensure the correct functioning of the equipment.
The main power plug socket and power switch are located on the back of the machine (Figure 27a).
To connect the iCELLis 500+ control system to the power supply, remove the socket protective cap by
turning the blue secure lock (Figure 27b-c). Confirm the control system is turned off, connect the power
plug firmly (Figure 27d) and secure the connector by turning the blue secure lock. Connect the power
cable to the appropriate power supply.