Purifiers are supplied with an over-temperature indicator, which
should be checked during installation. If the indicator turns
black, contact Pall Corporation SLS for technical assistance.
INSTALLATION AND SAFETY NOTES:
• Only trained personnel should install, operate, and maintain gas control
equipment.
• All compressed gases are potentially hazardous and must be handled
only by trained personnel. Incorrect execution of operations can cause
fire, explosion, atmospheric release of hazardous gas, or costly
contamination of process equipment.
• Safety Data Sheets (SDS) for all gases and purification materials used
in the facility should be available for consultation by all concerned
personnel. These data sheets are obtainable from gas suppliers and
Pall Corporation.
• An upstream process gas pressure regulator and pressure relief device
must be installed when a potential exists for overpressurization (
i.e.
pressure surge beyond MAWP of assembly).
• For in-line purifiers (PG550 and PG2400 models) the inlet fitting must
be installed finger-tight with inert gas purge flow introduced within
one minute after removal of the protective cap. Delay may result in
degradation of the purification material.
• For purifiers that are a part of a bypass manifold systems and the
PG11000, do not at any time prior to purifier installation attempt to
open manual valves or otherwise allow air to contact the purification
material bed. Valves must remain closed until installation is complete
and all equipment has been suitably leak tested and purged.
• Before starting a job, all personnel responsible for equipment installation
or operation must fully understand the specific procedures to be
accomplished and all pertinent safety considerations.