Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
7
PICO-155S Precision Saw
I
NSTRUCTION
M
ANUAL
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: [email protected] Web: http://www.metallographic.com
2.3.2 Installation - Micrometer
1. For easier installation, retract
micrometer adjustment arm - do
not force movement beyond limits
as this may damage the
micrometer.
2. Loosen bolts on sample support
arm and slide smaller diameter side
of micrometer shaft into outer shaft
guide. DO NOT tighten bolts on
sample support arm at this point.
3. Push micrometer into the outer shaft
guide thus pushing the micrometer
shaft into the machine. While
holding micrometer in place tighten
set screw on outer shaft guide.
NOTE: The hole in the micrometer
does not need to line up with the
bolt and DO NOT over tighten the
bolt as this may damage and pinch
the micrometer.
4. Attached sample fixture to sample
support arm and then align holder
to approximately where the
wafering blade will be installed.
Tighten sample holding arm bolts
Retract micrometer arm
Slide smaller end of
micrometer shaft into outer
shaft guide
Tighten set screw
- DO NOT OVER
TIGHTEN as this
may damage
micrometer
Install sample
fixture, align
sample support
arm and then
tighten bolts