4
O p e r a t i o n
5
5-1. Turning the power ON and OFF
1. Press the ON/OFF key to turn the power ON.
All the display symbols are displayed as shown above.
(About units: Only the units available illuminate.)
When the weighing value internally becomes stable, the display turns off except for a weighing unit and a
decimal point.
The balance waits for the weighing data to become stable, and zero will be shown with the ZERO mark
(power-on zero).
The range for power-on zero is within ±10% of the weighing capacity around the calibrated zero point.
If the power is switched ON while there is a load beyond this range, the balance is tared to zero and the
NET mark and the ZERO mark turn on.
2. Pressing the ON/OFF key again, and the power will be switched OFF.
• Auto-power off function
It is possible to have the power automatically switched OFF, if zero is displayed for
approximately 5 minutes. See “8-5. Function list” and set the function “
poff
”.
5-2. LCD backlight
The LCD backlight will turn on when the weight value changes more than 4 display digits or any key
operation is done. When the weight data becomes and stays stable for some time, it will automatically
turn off. There is also a setting that will make the backlight always stay on or off. For details, see the
function setting “
Ltup
” of “Function list”.
Содержание E12001