E
Series
User Manual
11
5 At step 3, pressing the PRINT key weighs the
zero-point value. Do not touch the pan during
weighing.
6 Place the calibration weight with the same
value as displayed on the pan. Press the
PRINT key to weigh it. Do not touch the pan
during weighing.
7
end
appears.
Remove the weight from the pan, and press
the CAL switch or MODE key to return to the
weighing mode.
• Note
The value set in step 4 is stored in memory
even after the power is switched off.
If the balance is to be moved to another
location, set the gravity acceleration value
for the current location and calibrate the
balance according to the procedure
above. See the next section to set the
value.
Weighing pan