10
This function adjusts the balance for accurate
weighing. Perform a calibration in the following
cases.
• When the balance is first used.
• When the balance has been moved.
• When the ambient environment has changed.
• For regular calibration.
6-1. Calibration using a weight
1 Warm up the balance for at least half an hour
with nothing on the pan.
2 Press and hold the calibration (CAL) switch until
Cal
appears, and release the switch.
3 The balance displays
Cal 0
.
To change the calibration weight value, proceed to
step 4.
To use the calibration weight value in the balance
memory, proceed to step 5.
4 Press the SAMPLE key. The display shows
the calibration weight value in "gram” that is
stored in the balance. Use the following keys to
change the value.
SAMPLE key To select the digit blinking to
change.
RE-ZERO key To set the value of the digit
selected.
PRINT key
To store the value and return
to step 3.
MODE key
To cancel the value and return
to step 3.
C a l i b r a t i o n
6
Press and hold
the CAL switch
Release the
CAL switch
Set the weight using
relevant keys
Calibration
(CAL) switch
Press and lower down the
calibration switch cover