-6-
6.3 Reverse Pipetting
This mode is generally used for pipetting high viscous liquids
accurately. It aspirates little excess liquid than the selected volume and
it dispense the exact liquid of the selected volume. The rest of the liquid
in the tip will need to be discarded.
Steps :
1. Select Reverse Pipetting from the applications.
2. Select volume by pressing up / down arrow key.
3. Press OK to finalize the volume.
4. Press trigger key to aspirate the volume.
5. Press trigger key to dispense the volume.
The excess liquid will remain in the tip.
6. Press trigger key to discard the excess liquid.
7. Press trigger key to repeat the process or press ESC key to
come out from the mode.
1. Select Multi Dispense from the applications.
2. Select Repeat Dispense from the Multi Dispense menu.
3. Select volume.
4. Select Dispensing Steps, the pipette will calculate the
maximum number of steps once you select the volume.
5. Use up/down arrow key to modify the dispensing steps and
press OK button.
6. Press trigger key to aspirate the liquid.
7. Press trigger key to dispense the liquid.
The excess liquid will remain in the tip.
8. Press trigger key to discard the excess liquid.
9. Press trigger key to repeat the process or press ESC key to
come out from the mode.
6.4 Multi Dispense
6.4.1 Repeat Dispense
The repeat dispense mode repeatedly dispenses the volume in
selected increments. For better accuracy, it will aspirate the excess
volume which will need to be discarded at the end.
6.4.2 Auto Dispense
Auto Dispense has a same function of Repeat Dispense but with an
automatic timer setting.
Steps :