3.5
Display Messages
The OxyLite front panel displays the pO
2
and temperature measurements
when a sensor is connected and in use. Other messages, including error
messages are also displayed to the user. A list of messages and their
meaning is provided below:
Message
Description
First message that is displayed during
the OxyLite boot sequence. Analogue
output during this time is 2.5V (50%
scale).
Second message displayed during the
boot sequence. Confirms instrument
model and internal software version.
Analogue output during this time is
2.5V (50% scale).
Third message of the boot sequence.
Displays (factory set) date and time
(GMT). Analogue output during this
time is 0V (0% scale).
There is no sensor connected to the
OxyLite but the monitor has completed
its boot sequence and is ready for use.
Analogue output during this time is
-2.5V (-50% scale).
First message displayed after a sensor
is connected. Confirms the sensor 6-
digit ID in the format XXnnnX, where X
is a letter and n is a number.
Second message after a sensor is
connected. Confirms hours of
accumulated longevity remaining and
the calibration period expiry date (see
section 3.7). If either limit has been
exceeded, press the SET button to
clear (see section 3.6).
Use Left nn hrs
Cal.Expiry dd/mm/yy
Loading
SensorID XXnnnX
No Sensor
Date dd/mm/yy
Time hh:mm
OxyLite
pO2_Soft 2.04
Oxford
Optronix