OxyLite User Manual
Revision 1.1 | May 2015
36
6.
T
ECHNICAL
S
PECIFICATIONS
Physical
Dimensions
95mm (H) x 290mm (W) x 260mm (D)
Weight
2Kg / 4.5lbs
Operating temperature
10 - 30°C
Operating humidity
0 – 70% (non-condensing)
Power requirements
VAC 100-240V, 50-60Hz, 30W max.
Fuse rating
2 x T1.6A anti-surge
Display
High contrast 40 character alphanumeric OLED
Analogue voltage outputs
2 x male BNC connectors, 2 outputs
LED excitation wavelength
525nm
Luminescence wavelength
650nm
Performance
Mode of operation
Luminescence decay lifetime (pO
2
) / T-type thermocouple
or manual user input (temperature)
Measurement units (displayed)
mmHg or kPa / ºC
Measurement range
0 – 200mmHg; 0 – 26.6kPa / 0 – 50ºC
Measurement resolution
0.1mmHg / 0.1ºC
Measurement accuracy
Oxygen:
±0.7mmHg (0 – 7mmHg)
±10% of reading (7 – 150mmHg)
±15% of reading (150 – 200mmHg)
Temp: ±0.2ºC
Measurement T90 response
time
< 20s (bare-fibre oxygen sensor) / < 2s (temperature)
Measurement sampling rate
1Hz
Measurement acquisition time
1s
Measurement time constant
(filtering)
5s (Oxygen); 1s (Temperature)
Display update time
2s (5s rolling averaged)
Validated temp compensation
range
10ºC – 45ºC
Sensor identification
Automatic; integrated connector EEPROM
Sensor calibration
Factory pre-calibration (unique per sensor); valid 6 months