Version 1.0 rev 14 Aug 2019
45
Balor
4.9.8 R
olling
s
HUtteR
- 100% d
Uty
c
ycle
e
xteRnal
e
xposURe
t
RiggeRing
In this mode, every positive edge of the external trigger will initiate a frame read out and start a new exposure. The
period of external trigger pulse defines exposure and cycle time for each frame read out.
On detection of the first positive edge, a frame read out is initiated, this frame is discarded as it does not contain the
correct exposure period. Reading out this first frame effectively begins the first exposure. When the next positive edge of
the external trigger is detected, a signal frame read out phase begins, ending the first exposure and starting the next.
Figure 18: Rolling Shutter - 100% Duty Cycle - External Exposure.
Table 9: Rolling Shutter - 100% Duty Cycle External Exposure Triggering.
Parameter
Minimum
Maximum
Exposure
1 Frame + 3 Row
3600 s
Cycle Time (EXT Trig Period)
Exposure
-
External Delay
1
2 Row
EXT Trig Pulse Width
2 Sensor Speed Clock Cycles
-
FIRE low period
1 Row
-