Version 1.0 rev 14 Aug 2019
40
Balor
Figure 12: Rolling Shutter Internal Triggering (Part 2: exposure time > readout time).
Figure 13: Rolling Shutter Internal Triggering (Part 3: Row 1, of the second exposure starts before, Row N, of the first exposure completes).
Table 5: Rolling Shutter Internal Triggering Timing Parameters.
Parameter
Minimum
Maximum
Exposure
1 Row
3600 s
Cycle Time (1/Frame Rate)
Maximum of: (Ex 4 Row) or (1
Frame + 2 Row)
20,000 s
* True minimum exposure time is the quoted value + T
res