Version 1.0 rev 14 Aug 2019
41
Balor
4.9.5 R
olling
s
HUtteR
- 100% d
Uty
c
ycle
i
nteRnal
t
RiggeRing
Rolling Shutter - 100% Duty Cycle in Internal Trigger maximizes the acquisition duty cycle and the frame rate by
removing the Reset phase from the acquisition sequence. In this mode, the action of reading out also starts the next
acquisition and hence the cycle time is the same as the exposure time.
Figure 14: Rolling Shutter - 100% Duty Cycle - Internal Triggering.
Table 6: Rolling Shutter - 100% Duty Cycle Internal Triggering Timing Parameters.
Parameter
Minimum
Maximum
Exposure
1 Frame + 1 Row
3600 s
Cycle Time (1/Frame Rate)
Exposure time
Exposure time
FIRE ANY low period
1 Row