48
Vulcan Series
Notice, do not expose the connectors to water and dust when the connector cover is open during
the cleaning process.
Cleaning the measurement window
Before starting the cleaning procedure, the device needs to be shut down and remove the battery
and disconnect from the mains.
The measurement window is recessed and located inside the instrument nose. To access the
measurement window, open the small hatch underneath the nose with a coin, screw driver or
similar. The nose will open and expose the measurement window and optional camera optics.
The window should be cleaned by using soft cotton buds. Pack of cotton buds are supplied with
the instrument. If the window is very dirty it can be cleaned by using optical grade isopropanol.
Do not use other detergents or solvents as they might leave a thin film on the measurement
window that might compromise the analysis results.
It is recommended to clean the measurement window every day before starting to take
measurements. This procedure guarantees the best possible results without the risk of window
contamination compromising the quality of the analysis results. When cleaning a very dirty
measurement window use optical grade isopropanol (isopropyl alcohol). Other solvents might
leave a film on the window surface that might compromise the measurement results.