Vulcan Series
21
Program types
There are two main types of measuring programs available, and both of them can be used with or
without the averaging feature.
●
Analyze
is the default measurement mode. It is used to identify unknown samples. After
measuring in Analyze mode, the closest matching grades are shown.
TIP
: You can customize
how many Good/Possible matches are shown in Grades settings.
●
Analyze with averaging
is an expanded version of normal Analyze. Averaging calculates the
average result from a preset amount of analyzes, which can be set when adding a Program.
If continuous averaging is selected instead of a preset amount, it can be stopped by tapping
the Done
.
●
Sort
is used when you are looking for one or more known grades and need to find out if
the sample belongs to these. It’s quick and shows immediately if the sample passes or
fails the sorting.
●
Sort with averaging
is similar to normal Sort, but with the averaging function as described
above.
See more with the Chemistry setting
Additional details about both Analyze and Sort results can be seen by switching on Chemistry by tapping
it’s icon
in the pull-down menu. It will display the concentrations of each element in the result and
it’s allowed minimum and maximum.
You can switch between a numerical and graphical view of the Chemistry data by tapping the numerical icon
or the graphical icon
.