HANDLING, INSTALLATION, STORAGE AND TRANSIT INFORMATION
KMW200CCD
Version 1.1
USER MANUAL
KMW200CCD Chiller v1.1
Page
13
M ains IN
CCD output
Outlet valve
Inlet valve and filter
Figure 5.2.1 Electrical and Water Connections on the Back and Side of the KMW200CCD
5.2.2.1 Equipment Required
20 litres of distilled water
Bucket
Funnel
Power cable for Oxford Diffraction CCD detector
Mains power cable for KMW200CCD
Water pipes (x2, thickness 10 x 3 mm)
5.2.2.2 Setting up Procedures
1. Connect the water pipes between the CCD detector and the KMW200CCD. Refer to figure 5.2.2
NOTE
: Water can flow through the Oxford Diffraction CCD detector in either direction.
Figure 5.2.2 Water pipe connection between KMW200CCD, Sapphire series and Onyx CCD
detectors
2. Remove the water fill cover located on the top of the KMW200CCD. Refer to figure 5.2.3.
Quick release fittings
Sapphire
Onyx
KMW200CCD
Pipes 10/3mm
Pipes 10/3mm