User’s Manual of
OWON
Colour Digital Storage Oscilloscope
61
The description of the
Auxiliary Function Menu
is shown as the following table.
Function Menu
Setting
Description
System Status
Display the system function menu.
Recall Factory
Call out the factory settings.
Do Self Cal
Carry out the self-calibration procedure.
Language
Chinese
English
Choose the display language of the
operating system.
Perform Self-Calibration
The self-calibration procedure can improve the accuracy of the oscilloscope under the
ambient temperature to the greatest extent. If the change of the ambient temperature is
up to or exceeds 5
℃
, the self-calibration procedure should be executed to obtain the
highest level of accuracy.
Before executing the self-calibration procedure, disconnect the probe or wire and the
input connector. Then
,
press the
F3
menu selection button and choose “
Perform
Self-Calibration
”. After confirming it is all set, push down
F3
button and choose
“
Perform Self-Calibration
”, entering the self-calibration procedure of the
instrument.
SYS STAT (System State)
Press the
F1
menu selection button and choose “
SYS STAT
” item. The menu pops up
in the screen as Fig.55.
Fig.55
SYS STAT Menu