50
Two trigger vertical point display on the screen when set for alternate trigger.
(red:CH1,yellow:CH2)
Alternate trigger menu display is shown as Fig40, Fig41:
Fig40
:
Alternate edge trigger menu
Fig41
:
Alternate video trigger menu
The set and edge trigger are the same when trigger type for edge. The set and video trigger
are the same when trigger type for video.
CH1 measurement 10K Hz sine wave, CH2 measurement a video signal. Alternate trigger
set steps are as follows:
1.
Press the
TRIG MENU
button and call out the trigger menu.
2.
Press the
F1
menu selection button and select
Alternate
for Type.
3.
Press the
F2
menu selection button and select source for
CH1
.
4.
Press the
F3
menu selection button and choose
Edge
for Type.
5.
Press the
F4
menu selection button and choose
Rising
for Slope.
6.
Press the
F5
menu selection button and choose next menu.
7.
Press the
F2
menu selection button and choose
DC
for Coupling.
8.
Press the F3 menu selection button and display 0.5div of Sensitivity CH1 trigger
complete set.
9.
Press the F4 menu button to select Holdoff ,use trig level knob to ajust Holdoff.
10.
Press the
F1
menu selection button and return to the first-level menu.
11.
Press F2 menu selection button and choose CH2 for source.