HDS2062M-N Handheld DSO & DMM
7-Advanced Function of Oscilloscope
Figure 35: DC Coupling
7.2.2 Open and Close Settings on Channel
With CH1 taken for example
.
Press
F2 Channel
key
first, and then
OFF
to
make a Close setting on CH1.
Press
F2 Channel
key
first, and then
ON
to
make an Open setting on CH1.
7.2.3 Setting the probe attenuation
To prevent excessive input voltage, we recommend you to set the probe attenuation level to the
10X position to prevent excessive voltage.
After attenuating the probe level by 10X, you also need to magnify the display level by 10X to
match the displayed amplitude with the real amplitude.
Press
F3 Probe
to adjust the probe attenuation level.
Table: Probe attenuation level and the corresponding menu setting
Probe attenuation level
Corresponding Menu Setting
1:1 1X
10:1 10X
100:1 100X
1000:1 1000X
7.2.4 Setting of Inverted Waveform
Inverted waveform: The displayed signal reverses 180 degrees relatively to the ground potential.
Press
F4 Invert
to
start Invert; press
F4 Invert
again to close Invert.
37