5.Front Panel Operation
31
(2)
Press
F2
to switch between Comma, Space, Off.
Take the Frequency parameter for instance:
Comma
Space
Off
To Set the Screen Saver
The screen saver will run automatically if no operation is taken for any key within the set
time. Press any button to resume.
(1)
Press
Utility
and choose
Disp Setup
, press
F3
to select Scrn Svr.
(2)
Press
F3
to switch between On/Off.
(3)
If On is selected, you can set the screen saver time. Turn the
knob
to change the
value; or press the number keys to input the desired time in minutes, press
F4
to
select the unit. The screen saver time range is 1 - 999 minutes.
To Set Output Parameter
To Set the Output Load
For the
Main output terminal
on the Front panel, the Generator has a built-
in 50Ωseries
impendence. If the actual load does not match the set one, the displayed amplitude and
offset are incorrect. This function is used to match the displayed voltage with the
expected one.
Steps for setting the Load:
(1)
Press
Utility
and choose
Output Setup
, press
F1
to select HighZ or *
Ω
( "*" represents a value).
(2)
To change the load value, after selecting *
Ω
, turn the
knob
to change it directly; or
press the number keys to input the desired value. Press
F3
or
F4
to select
the unit. The load range is 1
Ω
- 10K
Ω
.
Note
:
The series have
a fixed 50Ω Series Impendence. No matter what Value the set parameter
is, if the real load is different from the set one, the displayed voltage will not equal the real
voltage.