User Manual – Owandy-RX AC
Owandy-RX AC
radiographic system (Fig. 1) consists of:
1. Owandy-RX AC TIMER
The timer is the control panel used to manage the exposure times and to safely use the tubehead.
To make the exposure, the control button with safety key is available.
The timer can be connected to n° 2 ac tubeheads.
In case of alternate current tubeheads the technology of the timer is “self - compensating”:
depending on the line voltage fl uctuation, the microprocessor automatically modifi es the predetermined exposure time
ensuring a constant dose to the patient.
This technological expedient avoids the repetition of the exposure because of over/under exposure errors.
2. BRACKET
The horizontal bracket is available in 3 different lengths (110 cm, 80 cm, 40 cm) and represents the support for the
pantograph arm. Its shaft is fi xed in a dedicated section of the timer (top or bottom) and allows for 180° movement.
3. PANTOGRAPH ARM
Thanks to the new shape and new mechanisms of the positioning arm, it can be adjusted in height and depth in order
to precisely explore any spot in its reach.
It is made of light alloy with an ABS coating.
4. Owandy-RX AC TUBEHEAD
The intra-oral tubehead Owandy-RX AC is a monoblock type and its light alloy housing contains an airtight compartment.
The high voltage transformer, the X-ray tube and the expansion chamber are submerged in highly dielectric insulating
oil inside a light alloy container.
The expansion chamber guarantees an adequate compensation to oil expansion for the entire temperature range.
The X-ray tube is located in the back part of the container, allowing a source-skin distance 50% higher than traditional
structures.
5. CONE
The collimator cone or Beam Limiting Device represents the applied part of the device. Made of transparent polycarbonate,
or alternatively of lead-coated polycarbonate, it ensures:
- the correct distance between focal spot and skin
- dimension, direction and centering of X-ray beam
- the realization of different radiographic technique (biting and parallel technique).
During X-ray exposition, the collimator cone comes in contact with the skin of the patient.
Before each exam, it is necessary to apply to the cone a disposable protective cover designed to cover the end part of
the X-ray generator.
Such protection is useful to avoid crosscontamination (from patient to patient).