OWANDY RX - HUMAN VERSION
OWANDY RX - HUMAN VERSION
Page 32 of 55
USER MANUAL • Owandy-RX AC • 02/2020 • NXACEN160A
User Manual – Owandy-RX AC
PROGRAMMING DEFAULT EXPO-
SURE VALUES
6.1. PROGRAMMING DEFAULT EXPOSURE VALUES
!
WARNING
The 17 programmed exposure times cannot be modifi ed in the Owandy-RX AC radiographic system.
Meanwhile it is possible to customize the default exposure values.
!
WARNING
After customizing, the “Chart of default exposure values” (refer to Chapter 5) are not valid any more
.
To program the new exposure values press the following keys
PLEASE NOTE
The “repeat” function automatically sets in when the key is kept pressed, so the time shown on the display scrolls
faster.
To confi rm the new program check the LED key
LED
MEMO ON
indicates that it is possible to save the new default exposure value.
Press the button for 3s until the acoustic signal confi rms the new default exposure values have been saved.
LED
MEMO OFF
indicates that it is not possible to save the new default exposure value
PLEASE NOTE
It is not possible to save data when the “range of exposure fi eld” exceeds the program-
med exposure time limits. (refer to the example in the next page)