User Manual – Owandy-RX AC
Guidance and manufacturer’s declaration – electromagnetic immunity
Owandy-RX AC
is intended to be used in the electromagnetic environment specifi ed below.
The customer or
Owandy-RX AC
operator must ensure that the device is used in this type of environment.
Immunity test
CEI EN 60601 test level
Compliance level
Electromagnetic
environment
guidance
Electrostatic
discharge (ESD)
CEI EN 61000-4-2
+/- 8 kV contact
+/- 15 kV air
CEI EN 60601-1-2
Test level
Floors should be wood, concrete or
ceramic tile. If fl oors are covered
with synthetic material, the relative
humidity must be at least 30%
Electrical fast
transient/burst
CEI EN 61000-4-4
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
CEI EN 60601-1-2
Test level
Mains power quality should conform
to that of typical commercial or
hospital applications.
Surge
CEI EN 61000-4-5
+/- 1 kV differential mode
+/- 2 kV common mode
CEI EN 60601-1-2
Test level
Mains power quality should conform
to that of typical commercial or
hospital applications.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
CEI EN 61000-4-11
<5 % UT for 0.5 cycles
(>95 % dip in UT)
40 % UT for 5 cycles
(60 % dip in UT)
70 % U
T
(30 % dip in U
T
)
for 25 cycles (50Hz)
for 30 cycles (60Hz)
<5 % UT for 5 onds
(>95 % dip in UT)
CEI EN 60601-1-2
Test level
Mains power quality should conform
to that of typical commercial or
hospital applications. If the Owandy-
RX AC operator requires continued
operation even during mains power
outage, we recommend powering the
system using a UPS.
Mains frequency
(50/60 Hz)
magnetic fi eld
CEI EN 61000-4-8
30 A/m
CEI EN 60601-1-2
Test level
Power frequency magnetic fi elds
must be at the typical level of standard
mains for commercial or hospital use.
Note: Ut is the AC mains voltage prior to the application of the test level.