3.2 Conditions of use

The product was developed for everyday use and must not be used for unusual activities such as
extreme sports (free climbing, paragliding, etc.). Furthermore, the product must not be used for
the  operation  of  motor  vehicles,  heavy  equipment  (e.g.  construction  machines),  industrial
machines, firearms or motor-driven equipment.
The product is intended 

exclusively

 for use on 

one

 patient. Use of the product by another person

is not approved by the manufacturer.

3.3 Indications

For patients with unilateral and bilateral amputations.

For patients with transcarpal and transradial amputation.

3.4 Contraindications

For patients with transhumeral amputation.

For patients with partial hand amputation.

For patients with shoulder disarticulation.

For patients who are not able to generate an adequately separable pattern.

3.5 Qualification

Fitting a patient with the product may only be carried out by O&P professionals who have been
authorised  by  Ottobock  after  completion  of  a  corresponding  training  course.  The  O&P  profes­
sional also has to have the technical qualifications required for the alignment of a prosthesis with
all required settings and adjustments.
The user must be instructed about the use of the product by authorised, qualified personnel. 
The  "O&P  professional  mode"  of  the  adjustment  app  may  only  be  used  by  qualified  personnel,
therapists and nursing staff after participating in the relevant product training and obtaining certi­
fication for the application. Additional product training courses may become necessary to qualify
for app updates.

4 Safety

4.1 Explanation of warning symbols

WARNING

Warning regarding possible serious risks of accident or injury.

CAUTION

Warning regarding possible risks of accident or injury.

NOTICE

Warning regarding possible technical damage.

4.2 Structure of the safety instructions

WARNING

The heading describes the source and/or the type of hazard

The  introduction  describes  the  consequences  in  case  of  failure  to  observe  the  safety  instruc­
tions. Consequences are presented as follows if more than one consequence is possible:
>

E.g.: Consequence 1 in the event of failure to observe the hazard

>

E.g.: Consequence 2 in the event of failure to observe the hazard

This  symbol  identifies  activities/actions  that  must  be  observed/carried  out  in  order  to  avert
the hazard.

Содержание Myo Plus

Страница 1: ...Instructions for use qualified personnel 5 Myo Plus...

Страница 2: ...the instructions for use Angewandte Symbole Symbole Used Symboles utilis s Simboli utilizzati S mbolos utilizados S mbolos utilizados Gebruikte symbolen Symboler som anv nds Anvendte symboler Benytted...

Страница 3: ...3...

Страница 4: ...1 2 3 4 5 7 8 6 1 2 3 4 5 7 8 6 1 2 3 13E190 4...

Страница 5: ...ly put the product into use in accordance with the information contained in the accompanying documents supplied According to the manufacturer Otto Bock Healthcare Products GmbH the patient is the oper...

Страница 6: ...Hand DMC plus 8E38 6 8E39 6 Sensor Hand Speed 8E38 8 8E39 8 MyoHand VariPlus Speed 8E38 9 8E39 9 8E41 9 Transcarpal Hand DMC plus 8E44 6 System Electric Greifer DMC VariPlus 8E33 9 8E34 9 DMC VariPlu...

Страница 7: ...echnical qualifications required for the alignment of a prosthesis with all required settings and adjustments The user must be instructed about the use of the product by authorised qualified personnel...

Страница 8: ...ents in the section Scope of delivery and accessories WARNING Changes or modifications to the product made independently Injury due to faulty operation or malfunction of the product Have any changes o...

Страница 9: ...CAUTION Skin irritation due to inadequate cleaning of the product Skin irritation due to contact with soiled electrode domes Clean the product only as described in the section Cleaning and care see pa...

Страница 10: ...the electrode domes Applying thread locking compound forms an insulating layer that prevents transmission of the EMG signals NOTICE Failure to observe the system requirements for the installation of...

Страница 11: ...n Required tools Drill twist drill 5 mm The remote electrodes have to be positioned as follows 1 Mark the position of remote electrode 1 below the ulna and 6 7 cm distal of the olecranon 2 Extend the...

Страница 12: ...s and wiring can be protected against moisture and dirt with film or adhesive tape 7 4 Installing the Myo Plus TR The Myo Plus TR can be secured on the inner socket with adhesive tape or a hook and lo...

Страница 13: ...ing a connection to the coaxial plug The following steps are required to connect the Myo Plus TR to a coaxial plug 1 2 3 4 5 7 8 6 4 Connect the cables of the Myo Plus TR to the coaxial plug see fig 4...

Страница 14: ...on the supplied Bluetooth PIN card can be read with the mobile device requirement QR code reader and camera 8 1 System Requirements See the information in the Apple App Store or Google Play Store rega...

Страница 15: ...s app 6 1 Access online help 2 Display and name of the currently selected motion sequence 3 Connection to component has been established 4 Display of the selected screen 5 Display of the currently sel...

Страница 16: ...is 9 Use All electrode signals are displayed in the Myo Plus app via the Myo Plus TR The product is set up and configured via the Myo Plus app All information on use is found in the menu texts of the...

Страница 17: ...for medical devices This product meets the requirements of the 2014 53 EU directive The product meets the requirements of the RoHS Directive 2011 65 EU on the restriction of the use of certain hazardo...

Страница 18: ...d 2 this device must accept any interference including interference that may cause undesired operation of this device L utilisation de ce dispositif est autoris e seulement aux conditions suivantes 1...

Страница 19: ...voltage 6 V 11 1 V DC Current draw Max 25 mA Power supply 757B35 3 5 MyoEnergy Integral or 757B20 757B21 Energy Pack Lifetime 5 years Operating time MyoBock hand approx 14 hours bebionic hand approx 8...

Страница 20: ...indicat or Event Charge level above 50 illuminated in green and Charge level below 50 flashing alternating green and orange Charge level below 5 illuminated in orange 13 1 2 Beep signals The followin...

Страница 21: ...3 Symbols Used Type BF applied part Declaration of conformity according to the applicable European directives In some jurisdictions it is not permissible to dispose of these products with unsor ted ho...

Страница 22: ...devices is unlikely Harmonics according to IEC 61000 3 2 Not applicable power below 75 W Voltage fluctu ations flicker accord ing to IEC 61000 3 3 Product meets the requirements of the standard Electr...

Страница 23: ...0 to 390 TETRA 400 Pulse modu lation 18 Hz 1 8 0 3 27 450 430 to 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 1 8 0 3 28 710 745 780 704 to 787 LTE band 13 17 Pulse modu lation 217 Hz 0 2 0 3 9...

Страница 24: ...203 Otto Bock Healthcare Products GmbH Brehmstra e 16 1110 Wien Austria T 43 1 523 37 86 F 43 1 523 22 64 info austria ottobock com www ottobock com The product is covered by the following patents Pat...

Отзывы: