Error message
Error no.
Content
Cause
Possible measure
Memory Error
(None)
Unspecified memory
error in control mod
ules
e.g. defective
plug
connection,
cable
break
Defective
control
module
Contact
specialist
dealer
PM Memory Error
(None)
Memory error in con
troller
e.g. defective
plug
connection,
cable
break
Defective controller
Check all plug con
nections and cables
on controller
If error persists: con
tact specialist dealer
Bad Cable
(None)
Error on the commu
nication
cables
between
control
modules
e.g. defective
plug
connection,
cable
break
Check all plug con
nections and cables
on the control mod
ules (incl. controller)
If error persists: con
tact specialist dealer
Module Error
(None)
Error in control mod
ules
e.g. unspecified error
in displayed control
module
Displayed
control
module defective
Check all plug con
nections and cables
If error persists: con
tact specialist dealer
System Error
(None)
System error
Non-attributable error
in control modules
(primarily from third-
party providers)
Check all plug con
nections and cables;
disconnect/test
any
modules from third-
party providers in
succession if applic
able
If error persists: con
tact specialist dealer
SID Detached
(None)
Special control dis
connected
Special control dis
connected from LCD
monitor (sip and puff
control)
Special control con
nection to LCD mon
itor defective
Check all plug con
nections and cables
between special con
trol connection and
sip and puff control;
check programming
parameters
If error persists: con
tact specialist dealer
1E07
External on/off switch
disconnected
from
control panel
Switch Detached
1E08
External profile/mode
switch disconnected
from control panel
e.g. defective cable
connection to control
panel
Defective switch or
control panel
Check cable connec
tion and switching
function of the extern
al switch
If error persists: con
tact specialist dealer
Gone to Sleep
(None)
Control unit switches
to sleep mode
Extended
inactivity
while control unit is
switched on
Switch control unit
off/on
Charging
(None)
Battery charger is
connected
Battery charger is
connected to char
ging receptacle
Disconnect
battery
charger prior to driv
ing
JS Static Timeout
(None)
Joystick holding time
exceeded
Excessive
joystick
deflection
detected
(control unit stops
drive system to pre
vent potential dam
age to the motors)
Switch control unit
off/on
If
error
persists:
replace control panel
/ joystick; contact
specialist dealer