105
Beep
signal
Vibration
signal
Event
Resolution
3x short
3x short
Maintenance note:
E.g.: Maintenance interval has been
exceeded, temporary disruption of a
sensor signal
•
Use the Cockpit app to check
the next maintenance date for
the prosthesis (see page 88). If
the date is within the next
month, make a maintenance
appointment with the O&P pro
fessional. For this appointment,
the O&P professional has to be
provided with the battery char
ger and power supply in addi
tion to the prosthesis and tube
adapter.
•
Conduct the self-test again by
connecting/disconnecting the
battery charger.
•
If the beep is emitted again and
the maintenance date has not
been reached or exceeded, you
should visit the O&P profes
sional soon. If necessary, the
O&P professional will forward
the prosthesis to an authorised
Ottobock Service Center.
•
The product can be used
without restrictions. However,
vibration signals may not be
generated.
Battery charge level
During the charging process, the current battery charge level is indicated by the number of LEDs
lit on the side of the charger.
LEDs
0
1
2
3
4
5
Battery
charge level
0%-10%
10%-30%
30%-50%
50%-70%
70%-90%
>90%
16.3 Directives and manufacturer’s declaration
16.3.1 Electromagnetic environment
This product is designed for operation in the following electromagnetic environments:
•
Operation in a professional healthcare facility (e.g. hospital, etc.)
•
Operation in areas of home healthcare (e.g. use at home, use outdoors)
Observe the safety notices in the section "Information on proximity to certain areas" (see
page 67).
Table 1 – Electromagnetic emissions for all devices and systems
Directives and manufacturer's declaration – electromagnetic interference
The product 'Genium X3 - Bionic Prosthetic System 3B5-X3 / Genium X3 3B5-2' is designed for
operation in an electromagnetic environment as specified below. The customer or user of the
product 'Genium X3 - Bionic Prosthetic System 3B5-X3 / Genium X3 3B5-2' must ensure that
the device is operated in such an environment.