Beep signal
Vibration
signal
Additional action per
formed
Event
1x short
1x short
Weight taken off prosthetic
leg and leg kept still for 1
second
Switching to basic mode
(mode 1) carried out.
2x short
2x short
Weight taken off prosthetic
leg and leg kept still for 1
second
Switching to MyMode 1 (mode 2)
carried out.
3x short
3x short
Weight taken off prosthetic
leg and leg kept still for 1
second
Switching to MyMode 2 (mode 3)
carried out.
16.2.2 Warnings/error signals
Error during use
Beep signal
Vibration signal
Event
Required action
–
1x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen
erated)
Overheated hydrau
lics
Reduce activity.
–
3x long
Battery charge level
under 25%
Charge battery soon.
Remaining operating time
approx. 24 hours
–
5x long
Battery charge level
under 10%
Charge battery soon
Remaining operating time
approx. 6 hours
5x long
5x long repeated
every 60 seconds
Error of moderate
severity (see
page 93)
e.g. a sensor is not
operational
Walking possible with
restrictions. Note the
change in flexion resist
ance.
The product must be
inspected by an O&P
professional immediately.
10x long
10x long
Charge level 5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Charge the battery.
30x long
1x long, 1x short
repeated every
3 seconds
Severe error/indic
ation of safety
mode activation
(see page 93)
e.g. one or more
sensors are not oper
ational
Attempt to reset this error
by connecting/discon
necting the battery char
ger.
If the error persists, use
of the product is prohib
ited. The product must be
inspected by an O&P
professional immediately.