Ottobock | 13
>
Tools required:
711M64 Mounting Tool, 711M1 Mounting Tool (Alignment Rod)
and either 711M2 Mounting Plate or vice
►
Install the mounting tool on the alignment rod.
INFORMATION: The alignment rod must be clamped in a
vice, vertically for donning and horizontally for removal.
►
When using the mounting plate:
INFORMATION: Install the mounting plate on a suitable
surface and at optimal working height.
Attach the alignment rod to the mounting plate.
►
When using a vice:
Clamp the alignment rod into a vice at optimal working height.
►
Set the prosthetic hand to hand position "Neutral position"
(Neutral Mode) and lock the wrist.
►
Install the prosthetic hand on the mounting tool.
5.2 Don the prosthetic glove on the prosthetic hand
INFORMATION
►
Work quickly, adhering to the sequence of the instructions provided.
►
NOTICE! Never heat the prosthetic glove (e.g. with a blow drier or oven).
►
NOTICE! Do not overstretch the textile glove during the
donning process.
Roll up the textile glove, slide it over the fingers of the prosthet
ic hand and unroll it over the prosthetic hand.