3.6.3 Possible problem solutions
If the knee joint does not extend during the stance phase after stance phase flexion when used with the
Helix
3D
hip joint, this may be due to a number of different reasons.Therefore, please check the possible causes and, if
applicable, make the specified adaptations to achieve knee joint extension during the stance phase:
C-Leg / Genium: no extension during the stance phase
Cause
Possible solution
The stance phase resistance of the Helix
3D
Hip Joint is
set too high.
Reduce the stance phase resistance (Section 3.6.2).
The stride length, which results from the free swing
range and the swing phase resistance, is set too high.
Increase the swing phase resistance and/or reduce the free
swing range (Section 3.6.2).
The prosthetic alignment is not optimal.
Control and optimisation of the sagittal prosthetic alignment
under load (section 3.4).
Stance phase damping of the knee joint is too low.
The stance phase damping settings of the knee joint should
be optimised for sitting down and for stepoverstep climb
ing and descending of stairs and ramps.
If the CLeg®, in addition, does not switch to the swing phase mode at the end of the stance phase when using
the Helix
3D
Hip Joint, this can have the following causes:
C-Leg® does not switch to the swing phase
Cause
Possible solution
The forefoot load value is not reached during rollover.
Check the forefoot load during walking using CSoft and
correct, if necessary
CSoft Instructions for Use: 647G268).
If the Helix
3D
Hip Joint does not go into extension at heel strike, this can have several causes. Therefore, please
check the following possible causes and, if applicable, make the specified adaptations to achieve extension of
the Helix
3D
Hip Joint at heel strike:
Helix
3D
: no extension at heel strike
Cause
Possible solution
The stance phase resistance of the Helix
3D
Hip Joint is
set too high.
Reduce the stance phase resistance (Section 3.6.2).
The stride length, which results from the free swing
range and the swing phase resistance, is set too high.
Increase the swing phase resistance and/or reduce the free
swing range (Section 3.6.2).
The prosthetic alignment is not optimal.
Control and optimisation of the sagittal prosthetic alignment
under load (section 3.4).
Содержание 7E10 Helix 3D
Страница 23: ...Helix Ottobock 23...
Страница 177: ...Helix Ottobock 177...
Страница 221: ...Helix Ottobock 221...
Страница 243: ...Helix Ottobock 243...
Страница 247: ...Helix Ottobock 247 1 3 I H Helix3D Helix3D 5 2 1 4 Helix3D Ottobock 1 5 H Helix3D 1 5 1 H H Ottobock Helix3D...
Страница 248: ...248 Ottobock Helix H Ot tobock Helix3D H Helix3D C Leg Genium H ISO 15032 2 3 1 5 2 I Helix3D...
Страница 249: ...Helix Ottobock 249 1 5 3 I Ottobock H Ottobock Ottobock H H H Helix3D...
Страница 250: ...250 Ottobock Helix H I Helix3D Ottobock I Otto Bock HealthCare I Ottobock I 2 2 1 Helix3D...
Страница 252: ...252 Ottobock Helix 3 1 1 2 3 4 5 cm 4 5 6 cm 5 A B...
Страница 253: ...Helix Ottobock 253 3 2 I 5 2 y y 2 1 1 2 1 3 2 y y 2 4 5...
Страница 254: ...254 Ottobock Helix 3 2 1 Helix3D 1 7Z53 1 2 2 1 2 4G576 I 3 2 Helix3D 5 I 4G576 2 5 mm 3 4 5...
Страница 256: ...256 Ottobock Helix 3 4 H 1 2 L A S A R Assembly 3 8 2...
Страница 257: ...Helix Ottobock 257 30 mm 5 mm 35 mm 1 2 6 3 4 5 0 5 mm...
Страница 259: ...Helix Ottobock 259 3 5 L A S A R Posture 1 2 L A S A R Posture 10 mm 30 mm Helix3D 3 6 PU 3 8 3 7...
Страница 264: ...264 Ottobock Helix 3 9 H I Ottobock I ISO 15032 2 3 I Helix3D Ottobock 4 Helix3D 1...