MADSEN Xeta Test Settings
MADSEN Xeta is supplied with factory default settings. These settings have been selected as those most likely to be used
in the specific test functions. The settings are either test-specific, or apply to the general functions of MADSEN Xeta.
The system saves a copy of the test settings as 'system test settings' the first time a stimulus is presented to a new patient.
They are also part of the settings saved when a user setting is stored.
The system test settings will be used at power up or, if no user setting is activated, when at new patient is chosen.
If a user setting is chosen the test settings associated with that user is used when a new patient is chosen.
When you switch to a new test, the settings used in the test you are leaving will be saved and reused when you return to
that test.
The default settings can be loaded from
.
Test setting types
You can change the settings of MADSEN Xetain three ways:
•
•
•
All changes made to the test specific and general settings are permanent and will be used the next time you switch on
MADSEN Xeta.
6.1
Test-specific settings
When you select a specific test, you can change a number of settings relating specifically to this test.
1. Select the specific test, for example
AT
, and press
Setup
to change settings.
2. Use the
Left level
knob to scroll between test settings.
3. Use the
Right level
knob to scroll between setting values.
4. Use the
Frequency
knob to scroll between frequency settings.
5. Press
Setup
to leave the
Setup
function.
6. The settings are saved and will be used when next you use this test.
6.1.1
Auto Threshold test settings
This Automatic Threshold test adheres to the modified (ascending) Hughson & Westlake method.
•
To speed up the test progress, see the setting
.
•
To change settings, press
Setup
.
•
To return to the Auto Threshold test, see
Automatic Air Conduction testing
.
Default starting fre-
quency
1000 Hz.
You cannot change this value.
If the test cannot determine a threshold at this starting frequency, the entire test is
aborted. Reinstruct the patient and try again.
AT Frequencies
•
Use the
Frequency
knob to scroll between frequencies.
•
Use the
Right level
knob to set to
On
or
Off
. 1000 Hz is always
On
and cannot
be changed.
Otometrics - MADSEN Xeta
55
Содержание Madsen Xeta
Страница 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Страница 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Страница 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Страница 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Страница 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Страница 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Страница 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Страница 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Страница 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Страница 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...