AS AT Procedure
If
AS Do AT when AS fails
is set to
Yes
, this setting also applies to the Auto
Threshold sequence.
•
Standard
The stimulus but not the preset stimulus duration is interrupted by the patient
response. This is followed by a random inter-stimulus interval.
•
Quick
Both the stimulus and the preset stimulus duration are interrupted by the
patient response. The response is immediately followed by the random inter-
stimulus interval.
This results in a faster test.
AS Discreet mode
•
On
The Signal indicator does not light up and Xeta does not sound a beep when
stimulus is presented and the stimulus levels are not displayed.
•
Off
The stimulus level is displayed and the signal indicator lights up when stimulus
is presented.
6.2
General settings
These settings apply to all test functions. However, some settings do not apply in automatic testing.
Changing general settings
1. To access the Setup menu of Xeta, press
Setup
.
When the
Setup
function is active, the
Setup
button flashes. Most other buttons are deactivated and you cannot run
any tests.
2. Turn the
Left level
knob to scroll through the parameter list.
As a rule, the top line in the display contains a description of the current parameter, for instance:
AC TRANSDUCER
The bottom line contains the value that you can change.
3. To change the setting, for most items, use the
Right level
knob.
4. Changes made under
Setup
will be saved when you leave
Setup
, and will apply the next time you switch on Xeta.
5. The most recently selected item will also be the one displayed the next time you activate
Setup
.
6. To exit this function press
Setup
.
6.2.1
Return to default settings
To return to the default settings, see
58
Otometrics - MADSEN Xeta
6 MADSEN Xeta Test Settings
Содержание Madsen Xeta
Страница 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Страница 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Страница 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Страница 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Страница 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Страница 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Страница 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Страница 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Страница 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Страница 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...