9.1.4
Laser
M
odulation
C
ommands
(
LM
)
cmd
type
min
max
default
unit
description
LMDI
bool
S
R
S
internal digital modulation
LMDX
bool
S
R
S
external digital modulation
LMAX
bool
S
R
S
external analog modulation
LMW
float
100
1000000
1000
µ
s
pulse width
LMP
float
LMW
+
100 600000000
2000
µ
s
pulse period
LMDIC
word
0
65534
0
number of pulses
LMDIO
word
0
65534
0
number of suppressed pulses
LMDXN
bool
R
S
S
negate modulation input
9.1.5
Laser
S
equencer
C
ommands
(
LZ
)
The
LZTR
command
is
available
in
every
laser
driver:
cmd
type
min
max
default
unit
description
LZTR
float
300
34000
300
ms
ramp time (refers to
I
max
)
The following commands are optionally available on request:
cmd
type
min
max
default
unit
description
LZR
bool
— no parameter —
ms
sequencer run (stop with
LS
)
LZP
word
ms
sequencer point select
LZPT
word
ms
subsequence time (duration)
LZPC
float
ms
subsequence current (end)
9.1.6
Pilot
L
aser
C
ommands
(
P
)
cmd
type
min
max
default
unit
description
PL
bool
S
R
S
pilot laser stop/run
PP
word
0
16
0
pilot laser modulation
9.2 Temperature Sensor and TEC Commands (
xT)
In
these
commands
x
has
to
be
replaced
by
a
digit
(or
letter)
to
select
the
temperature
sensor
or
TEC.
The
first
temperature
sensor
or
TEC
corresponds
to
1
,
the
second
to
2
,
the
third
to
3
and
so
on.
31
Содержание Laser Lab Source LDC Series
Страница 1: ...Operating Manual LDI LDC and TEC Series Laser Diode Drivers and TEC Controllers...
Страница 2: ......
Страница 34: ...0x0040 LMDX ON 0x4000 gate option 0x0080 LMAX ON 0x8000 reduced mode 34...
Страница 35: ......
Страница 36: ...Copyright 2009 2020 OsTech e K Version 1 2 2016 06...