8
.
Error Codes
When
an
error
occurs,
the
driver
generates
an
error
code.
The
error
code
can
be
determined
by
the
GE
command.
Furthermore,
the
error
is
shown
on
the
display.
A
list
of
error
codes
and
their
causes
follows:
error code
cause
0
no
error
1
interlock open
2
laser
compliance
voltage
not
acceptable
or
no
laser
connected
3
internal
supply
voltage
not
acceptable
4
laser temperature sensor open
5
crystal temperature sensor open
6
laser temperature exceeds upper limit
7
laser temperature lower than lower limit
8
laser short-circuit or no laser connected
9
device temperature (
GT
) too high
10
laser temperature exceeds maximum laser temperature (
LTM
)
11
crystal temperature exceeds upper limit
12
crystal temperature lower than lower limit
16
laser current greater than maximum current limit (
LCLM
)
17
current error
18
total power limit exceeded
29