background image

16

Electromagnetic Immunity Compliance Information

Immunity Test

Standard

Compliance Level

Electrostatic Discharge

IEC61000-4-2

± 8 kV contact

± 15 kV air

Electrical Fast Transient/Burst

IEC61000-4-4

± 2 kV for power supply lines

± 1 kV for input/output lines

Surge

IEC 61000-4-5

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

Voltage Dips

IEC61000-4-11

0 % UT; 0.5 cycle  

At 0°, 45°, 90°, 135°, 180°, 225°, 270° 

and 315° 

0 % UT; 1 cycle and  

70 % UT; 25/30 cycles  

Single phase: at 0°

Voltage Interruptions

IEC61000-4-11

0 % UT; 250/300 cycle 

Power Frequency (50/60 Hz) Magnetic Field

IEC61000-4-8

30 A/m

Conducted Disturbances Induced by RF Fields

IEC 61000-4-6

3 V   

0.15 MHz – 80 MHz  

6 V in ISM and amateur bands  

between 0.15 MHz and 80 MHz   

80 % AM at 1 kHz   

Radiated RF EM Fields

IEC 61000-4-3

10 V/m  

80 MHz – 2.7 GHz 

80 % AM at 1 kHz

Содержание Power Knee

Страница 1: ...POWER KNEE Instruction for Use...

Страница 2: ...USE 7 OPERATION 9 TECHNICAL DATA 12 MAINTENANCE 13 STORAGE TRANSPORT CONDITIONS 13 PARTS ORDERING 13 WARRANTY 13 DISPOSAL OF EQUIPMENT 13 PRODUCT CERTIFICATION INFORMATION 13 BLUETOOTH MODULES REGULAT...

Страница 3: ...EDs C B A Rear of prosthesis On Off switch Prosthesis status LED Front of prosthesis Battery Lock tab 1 2 Front Back Battery status button On Off switch Prosthesis status LED Battery status LEDs 4 5 6...

Страница 4: ...Front Battery Fan Status LED Battery charger Battery charger Power inlet Power supply Wall outlet Charging Flashes Charge completed Steady Problem Battery charger status LED OK Prosthesis Status LED P...

Страница 5: ...action INTENDED USE POWER KNEE is designed for transfemoral and knee disarticulation amputees with moderate to high activity level In addition POWER KNEE can be used for an extended user population na...

Страница 6: ...nd manner To protect natural resources and to promote material reuse please separate batteries and electronic components from other types of waste and recycle them through your local free electronic p...

Страница 7: ...ures between 10 C and 40 C Caution Power on the prosthesis only when you are wearing it Caution Follow recommended maintenance intervals INDICATIONS FOR USE The POWER KNEE is designed for patients wit...

Страница 8: ...escending beep is heard and the prosthesis status LED goes off A Caution Never remove the battery if the prosthesis is not powered off as data could be lost 3 Push down the lock tab as shown in B belo...

Страница 9: ...adjusts its behavior to activities like walking ascending or descending stairs ramps etc When none of these activities are detected it operates in standing state allowing you to move effortless in con...

Страница 10: ...descending ramps and stairs KNEELING 1 Allow the prosthetic knee to reach the standing state 2 Place and maintain load onto the prosthesis 3 Put your sound leg in front of the prosthesis 4 Force the p...

Страница 11: ...tivity feel the prosthesis temperature with your hand to confirm if this is the cause of the problem and let the prosthesis cool before resuming your current activity The risk of an overheat warning m...

Страница 12: ...3 15 16 x 3 15 16 Weight 2 7 kg 6 lb w o battery Rating 50 4 VDC 25 A max Angle of knee flexion Max 120 Expected lifetime 5 years Foot Module Variflex LP Variflex Talux Variflex XC Re flex ShockTM Re...

Страница 13: ...ANTY For warranty information see the POWER KNEE Warranty Card DISPOSAL OF EQUIPMENT Dispose of the POWER KNEE accessories in an environmentally responsible manner Never incinerate a battery as it may...

Страница 14: ...rticular installation If this equipment does cause harmful interference to radio or television reception which can be determined by tuning the equipment off and on the user is encouraged to try and co...

Страница 15: ...d result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Portable RF communications equipment including peripherals such...

Страница 16: ...s 2 kV line s to earth Voltage Dips IEC61000 4 11 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles Single phase at 0 Voltage Interruptions IEC61000 4 11 0 UT 250 3...

Страница 17: ...8 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 19...

Страница 18: ...with other recommended ssur components If un usual movement or product wear is detected in a structural part of a device at any time the patient should be instructed to immediately discontinue use of...

Страница 19: ...1 8152 Alcobendas Fax 86 21 6127 1799 Fax 1 866 441 3880 Madrid Espa a asia ossur com Tel 00 800 3539 3668 ssur Europe BV Fax 00 800 3539 3299 ssur Australia De Schakel 70 orders spain ossur com 26 Ro...

Отзывы: