guidance and manufacturer’s declaration – electromagnetic emissions

The device is intended for use in the electromagnetic environment specified below. The customer or user of device should 

assure that it is used in such an environment.

emissions Test

compliance

electromagnetic environment – guidance

RF Emissions CISPR 11

Group 1

The device uses RF energy only for its internal function. Therefore, its RF 

emissions are low and are not likely to cause any interference in nearby electronic 

equipment.

RF Emissions CISPR 11

Class B

The device is suitable for use in all establishments, including domestic 

establishments and those directly connected to the public low voltage power 

supply network that supplies buildings used for domestic purposes.

Harmonics Emissions

IEC 61000-3-2

Class B

Voltage Fluctuations/

Flicker Emissions 

vIEC 61000-3-3

Complies

guidance and manufacturer’s declaration – electromagnetic emissions

The device is intended for use in the electromagnetic environment specified below. The customer or user of device should 

assure that it is used in such an environment.

immunity Test

iec 60601 Test level

compliance level

electromagnetic environment – 

guidance

Electrostatic Discharge 

(ESD)

IEC 61000-4-2

±6 kV  contact

±8 kV  air

±6 kV  contact

±8 kV  air

Floors should be wood, concrete, 

or ceramic tile. If floors are 

covered with synthetic material, 

the relative humidity should be at 

least 30%.

Electrical fast 

transient/burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

±2 kV for power supply lines

±1 kV for input/output lines

Mains power quality should be 

that of a typical commercial or 

hospital environment.

Surge IEC 61000-4-5

±1 kV line(s) to line(s)

±2 kV line(s) to earth

±1 kV line(s) to line(s)

±2 kV line(s) to earth

Mains power quality should be 

that of a typical commercial or 

hospital environment.

Voltage dips, Short

interruptions and 

voltage variations on 

power supply

lines IEC 61000-4-11

<5 % U

T

 (>95 % dip in U

T

) for 

0.5 cycle.

40 % U

T

 (60 % dip in U

T

) for 5 

cycles.

70 % U

T

 (30 % dip in U

T

) for 

25 cycles.

<5 % U

T

 (>95% dip in U

T

) for 

5 s.

<5 % U

T

 (>95 % dip in U

T

for 0.5 cycle.

40 % U

T

 (60 % dip in U

T

) for 

5 cycles.

70 % U

T

 (30 % dip in U

T

) for 

25 cycles.

<5 % U

T

 (>95% dip in U

T

) for 

5 s.

Mains power quality should be 

that of a typical commercial or 

hospital environment. If the user 

of the device requires continued 

operation during power mains 

interruptions, it is recommended 

that the device be powered from 

an uninterruptible power supply 

or a battery.

Power frequency 

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields 

should be at levels characteristic 

of a typical location in a typical 

commercial or hospital 

environment.

Note: U

T

 is the a.c. mains voltage prior to application of the test level.

Содержание PROPRIO FOOT

Страница 1: ...PROPRIO FOOT Technical Manual...

Страница 2: ...uction 7 Technical specifications 7 Setup 8 Basic operation 11 Calibration 13 Functions 14 Advanced operation 15 Maintenance 16 Training tips 17 Warranty 17 Electromagnetic compatibility precautions 1...

Страница 3: ...3 1 3 1 3 1 3 1 2 1 2 1 2 1 2 1 2 1 2 a b c d e f...

Страница 4: ...2 1 start 1 move 1 complete Move to neutral relax 4 1 relax Manual motor movement 6 n a Sound on off 7 1 on calibration on off 9 1 on off Default manufacturer setting Stair Mode 4 Stair mode Sequence...

Страница 5: ...and should never be disposed of in residential or commercial garbage They should never be incinerated because they may explode Perform field servicing at the recommended intervals according to the in...

Страница 6: ...mpanying documents Consult Instructions for use Meets IEC type B leakage current requirements CE label with Notified Body indentification number Must be disposed of or recycled properly Connection for...

Страница 7: ...le environmental conditions are described in the technical specifications see Technical specifications proprio foot is suitable for continuous use The proprio foot may only be fitted by a qualified an...

Страница 8: ...attery in the most proximal position possible Avoid flexion limitations to the knee Check the position using the dummy The position of the battery assembly may vary on individual sockets Care must be...

Страница 9: ...ion during walking and make changes if necessary to the assembly BENCH ALIGNMENT Figure F Overview Fit the foot with the selected cover and shoe Power ON see chapter Power ON Adjust the Heel Height to...

Страница 10: ...ely note do not use a Flex Foot shoe horn or any other lever to remove the foot cover as this may cause damage to the ankle unit CHARGING AND POWER Figure K Only use the power supply supplied for prop...

Страница 11: ...d will be disabled after 60 seconds from use To enable the keypad hold both buttons at least for one second When proprio foot is activated for the first time the default manufacturer settings will be...

Страница 12: ...te If the heel height exceeds 50mm 2 a signal will be heard and the unit will not adjust to the heel height Change shoes for lower heel height and repeat the adjustment Ensure free ankle motion Heel h...

Страница 13: ...d until completion of calibration is confirmed by a beep and vibration The LED blinks green The system is now properly calibrated to the user and enabled for normal adaptive functions note Make sure t...

Страница 14: ...lting in unexpected ankle adaptation STAIRS During stair ascent proprio foot will raise the toe This will facilitate stair ambulation and improve gait symmetry When leading with the prosthetic side pr...

Страница 15: ...Mode the proprio foot will move into CHAIR EXIT MODE To initiate this use one of the following methods 1 Tap the heel of the foot 2 Flex the knee and place the foot beneath the chair Lift the foot to...

Страница 16: ...m confirms with a beep and vibration when enabled There is no confirmation when sound is disabled MAINTENANCE SERVICE proprio foot is designed and manufactured to provide long and trouble free service...

Страница 17: ...rosthetic side upstairs and to lead with the prosthetic side when going down This will provide stair adaptation after the first step on stairs Note Instruct the user to be careful walking downstairs f...

Страница 18: ...air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2kV for powe...

Страница 19: ...RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the fol...

Страница 20: ...distance between portable and mobile RF communications equipment transmitters and the device as recommended below according to the maximum output power of the communications equipment Rated maximum ou...

Страница 21: ...evel Category ssur high activity Category Weight Kg Lable text 1 52 P3 ISO 10328 P3 52 kg 2 59 P3 ISO 10328 P3 59 kg 3 68 P4 ISO 10328 P4 68 kg 4 77 P4 ISO 10328 P4 77 kg 5 88 P4 ISO 10328 P5 88 kg 6...

Страница 22: ......

Страница 23: ...with other recommended ssur components If un usual movement or product wear is detected in a structural part of a device at any time the patient should be instructed to immediately discontinue use of...

Страница 24: ...6127 1799 Fax 1 866 441 3880 Madrid Espa a asia ossur com Tel 00 800 3539 3668 ssur Europe BV Fax 00 800 3539 3299 ssur Australia De Schakel 70 orders spain ossur com 26 Ross Street 5651 GH Eindhoven...

Отзывы: