The SmartPeg is available with different connection geometries to 

fit all major implant products on the market. You can find all 

 

available SmartPegs on 

osstell.com/smartpegguide.

The SmartPegs™ are disposable and should only be used for one 

or multiple measurement at one treatment session, for a single 

patient only (to avoid cross-contamination). Repeated re-use may 

result in false readings due to wear and tear of the soft aluminium 

SmartPeg threads.

11) How to Measure

Prior to use on a patient, place a barrier sleeve over the Osstell 

Beacon. The barrier sleeve helps prevent cross-contamination and 

helps keep dental composite material from adhering to the surface 

of the instrument tip and body, and discoloration and degradations 

from cleaning solutions.

Note: 

•  Barrier sleeves are single patient use only.

•  Discard used barrier sleeves in standard waste after each patient.

•  Do not leave barrier sleeves on the instrument for extended periods.

•  See below for recommended barrier sleeves.

  TIDIshield, Art no: 21021, Art no: 20987. 

www.tidiproducts.com

  PremiumPlus: 123, Small short 123, Small

  Please also see additional recommended barrier sleeves on:  

osstell.com/get-started-beacon

• The Osstell Beacon instrument must be cleaned and sanitized 

with appropriate cleaning and/or sanitizing agents after each 

patient. See section 15) Cleaning and Maintenance for acceptable 

agents.

A first measurement should be taken at implant placement to get a 

baseline for future measurements throughout the healing process. 

Before the final restoration, another measurement is taken which 

makes it possible to observe the stability development of the 

implant.

It is recommended to measure in both Buccal-Lingual and Mesial- 

Distal direction to find the lowest stability. Therefore, the Osstell 

Beacon prompts the user to measure in both these directions.

We recommend you studying the more detailed information (videos 

and quick guides) available on 

osstell.com/get-started-beacon,

 

to utilize the full functionality of your Osstell Beacon.
1. Activate the Osstell Beacon by picking it up. The instrument will 

start-up and after showing the battery status, the instrument will 

be ready for measurement in the BL (Buccal – Lingual) direction, 

which is prompted in the upper display as well as optimal angle 

of the instrument tip towards the Smartpeg. 

2. Place a barrier sleeve over the Osstell Beacon instrument. 

See fig 1.

3. Place the SmartPeg into the SmartPeg Mount. The SmartPeg is 

magnetic, and the SmartPeg Mount will hold the SmartPeg. 

See fig 2. 

Attach the SmartPeg to the implant or abutment by screwing the 

SmartPeg Mount using finger force of approximately 4-6 Ncm. 

 

Do not over-tighten, to avoid damaging the SmartPeg threads.

4. Hold the instrument tip close (3-5mm) to the top of the SmartPeg 

without touching it. An audible sound indicates when measure-

ment has started, and measured data will be shown in the upper 

display combined with a colored light indication below the 

instrument tip. Bring the instrument out of the mouth to clearly 

read the ISQ value and the colored indication. 

Содержание Beacon

Страница 1: ...osstell com USER MANUAL Osstell Beacon GB ...

Страница 2: ...rm a measurement in two directions Buccal lingual and Mesial Distal as guided by the instrument This is impor tant to detect the lowest implant stability The SmartPegs are disposable and should only be used for one or multiple measurements at one treatment session for a single patient only to avoid cross contamination Repeated re use may result in false readings due to wear and tear of the soft al...

Страница 3: ...verall treatment evaluation program The final implant treatment decisions are the responsibility of the clinician Contraindications Osstell Beacon is contraindicated for implant systems to which the SmartPeg could not be attached for mechanical incompatibility reasons See section 10 for more information about SmartPegs Osstell Beacon is contraindicated when used together with Pegs not approved by ...

Страница 4: ...lectronic product is approved by Federal Communications Comission FCC FCC ID QOQBLE113 IC 5123A BGTBLE113 Non ionizing electromagnetic radiation Not Sterilizable Temperature limitation Humidity limitation Atmospheric pressure limitation 6 Safety Symbols 1 2 3 4 5 6 Your Osstell Beacon system includes the following items 1 Osstell Beacon Instrument 2 Osstell USB cable 3 Osstell SmartPeg Mount 4 Oss...

Страница 5: ...acon Instrument The Osstell Beacon is activated by a built in motion sensor As soon as the motion sensor detects movement the Osstell Beacon will start up flash red yellow green and shortly show battery status in the lower display and then be ready for measurement in the BL Buccal Lingual direction which is indicated in the upper display The Osstell Beacon will start to measure indicated by an aud...

Страница 6: ...implant placement to get a baseline for future measurements throughout the healing process Before the final restoration another measurement is taken which makes it possible to observe the stability development of the implant It is recommended to measure in both Buccal Lingual and Mesial Distal direction to find the lowest stability Therefore the Osstell Beacon prompts the user to measure in both t...

Страница 7: ...ould be taken on the implant before the abutment is attached and then a second measurement on the abutment 13 Interpret the Result Implant Stability An implant can have different stability in different directions The stability of the implant is dependent of the surrounding bone configuration There is often a direction where the stability is lowest and a direction where the stability is highest and...

Страница 8: ...ze or soft cloth with an acceptable see list below surface disinfectant and wipe the whole Osstell Beacon instrument Note Do not autoclave the Osstell Beacon instrument Routinely check the surface of the instrument tip and overall surface for possible cracks and residuals Acceptable Cleaners Isopropyl alcohol Ethyl alcohol based cleaners Do not use Strongalkalidetergentofanytype includinghandsoaps...

Страница 9: ...mbols used follow the European standard EN 60601 1 and ISO standards 9687 and 15223 as far as possible Notes on electromagnetic compatibility EMC Medical electrical equipment is subjected to particular precautions with regards to EMC and must be put into operation in accordance with the EMC notes included below Osstell guarantees the compliance of the device with the EMC requirements only when use...

Страница 10: ... to 1060 hPa IP class IP20 17 Troubleshooting No measurement or unexpected value Re Used SmartPeg The SmartPegs are disposable and should only be used for one or multiple measurement at one treatment session for a single patient only Repeated re use may result in false readings due to wear and tear of the soft aluminium SmartPeg threads Wrong SmartPeg type selected for the Implant See SmartPeg ref...

Страница 11: ...t the local sales representative or distributor for further instructions Difficult to measure in an exact recommended directionn No space e g due to adjacent teeth Try to measure at a slightly different angle Difficulties attaching the SmartPeg Wrong SmartPeg Ensure that the SmartPeg is compatible with the implant system 18 Service and Support In the event of an instrument malfunction contact the ...

Страница 12: ...Osstell AB Stampgatan 14 411 01 Göteborg Sweden Phone 46 31 340 82 50 info osstell com www osstell com Osstell AB Date of issue 2018 05 25 25087 00 ML OSSTELL Beacon USER MANUAL ...

Отзывы: