25087-00 ML

5

IP20

Protected against solid foreign objects of 12.5mm 

∅ 

and greater. No protection against water. 

Use by date

Lot/batch code

Sterilized using irradiation

RX-

Only

For US market only: Prescription use only. U.S. Federal 

Law restricts this device to sale by or on the order of a 

licensed practitioner

REF

Catalog Number

2

Do not reuse

7) Before You Start

 

The Osstell Beacon is delivered from factory in a “transport” mode,  

where the built-in motion sensor is deactivated.

To de-activate the transport mode and start charging, connect the 

small USB connector of the Osstell USB cable to wide end of the 

instrument. Connect the large USB connector to a standard USB 

type A outlet of a PC, laptop or charger.

The Osstell Beacon will start-up and enter charging mode. Charge 

the Osstell Beacon for at least 3 hours or until the Osstell Beacon 

indicates full charge. Remove the Osstell USB cable.

Note! It is not possible to perform an ISQ measurement during

charging.

8) Operation of the Osstell Beacon  

Instrument

The Osstell Beacon is activated by a built-in motion sensor. As soon 

as the motion sensor detects movement, the Osstell Beacon will 

start-up, flash red-yellow-green and shortly show battery status 

in the lower display and then be ready for measurement in the BL 

(Buccal – Lingual) direction, which is indicated in the upper display.
The Osstell Beacon will start to measure, indicated by an audible 

sound, when near an Osstell Smartpeg or Testpeg due to a built-in 

magnetic detector that senses the magnet on the top of the Osstell 

Smartpeg/TestPeg. Measured data will be shown in the upper display 

combined with a colored light indication below the instrument tip.
The red, yellow and green color represents low, medium or high 

implant stability. After 60 seconds of no motion, the Osstell Beacon 

will turn off automatically.

9) TestPeg

The Osstell TestPeg may be used for testing and learning how to  

use the system. Place the Osstell TestPeg on the table or hold it in 

your hand. Activate the Osstell Beacon by moving it and hold the  

instrument tip approximately 3-5mm away from the top of the 

Osstell TestPeg. The Osstell Beacon should start to measure and 

present an ISQ value of 55 +/- 2 ISQ.

Содержание Beacon

Страница 1: ...osstell com USER MANUAL Osstell Beacon GB ...

Страница 2: ...rm a measurement in two directions Buccal lingual and Mesial Distal as guided by the instrument This is impor tant to detect the lowest implant stability The SmartPegs are disposable and should only be used for one or multiple measurements at one treatment session for a single patient only to avoid cross contamination Repeated re use may result in false readings due to wear and tear of the soft al...

Страница 3: ...verall treatment evaluation program The final implant treatment decisions are the responsibility of the clinician Contraindications Osstell Beacon is contraindicated for implant systems to which the SmartPeg could not be attached for mechanical incompatibility reasons See section 10 for more information about SmartPegs Osstell Beacon is contraindicated when used together with Pegs not approved by ...

Страница 4: ...lectronic product is approved by Federal Communications Comission FCC FCC ID QOQBLE113 IC 5123A BGTBLE113 Non ionizing electromagnetic radiation Not Sterilizable Temperature limitation Humidity limitation Atmospheric pressure limitation 6 Safety Symbols 1 2 3 4 5 6 Your Osstell Beacon system includes the following items 1 Osstell Beacon Instrument 2 Osstell USB cable 3 Osstell SmartPeg Mount 4 Oss...

Страница 5: ...acon Instrument The Osstell Beacon is activated by a built in motion sensor As soon as the motion sensor detects movement the Osstell Beacon will start up flash red yellow green and shortly show battery status in the lower display and then be ready for measurement in the BL Buccal Lingual direction which is indicated in the upper display The Osstell Beacon will start to measure indicated by an aud...

Страница 6: ...implant placement to get a baseline for future measurements throughout the healing process Before the final restoration another measurement is taken which makes it possible to observe the stability development of the implant It is recommended to measure in both Buccal Lingual and Mesial Distal direction to find the lowest stability Therefore the Osstell Beacon prompts the user to measure in both t...

Страница 7: ...ould be taken on the implant before the abutment is attached and then a second measurement on the abutment 13 Interpret the Result Implant Stability An implant can have different stability in different directions The stability of the implant is dependent of the surrounding bone configuration There is often a direction where the stability is lowest and a direction where the stability is highest and...

Страница 8: ...ze or soft cloth with an acceptable see list below surface disinfectant and wipe the whole Osstell Beacon instrument Note Do not autoclave the Osstell Beacon instrument Routinely check the surface of the instrument tip and overall surface for possible cracks and residuals Acceptable Cleaners Isopropyl alcohol Ethyl alcohol based cleaners Do not use Strongalkalidetergentofanytype includinghandsoaps...

Страница 9: ...mbols used follow the European standard EN 60601 1 and ISO standards 9687 and 15223 as far as possible Notes on electromagnetic compatibility EMC Medical electrical equipment is subjected to particular precautions with regards to EMC and must be put into operation in accordance with the EMC notes included below Osstell guarantees the compliance of the device with the EMC requirements only when use...

Страница 10: ... to 1060 hPa IP class IP20 17 Troubleshooting No measurement or unexpected value Re Used SmartPeg The SmartPegs are disposable and should only be used for one or multiple measurement at one treatment session for a single patient only Repeated re use may result in false readings due to wear and tear of the soft aluminium SmartPeg threads Wrong SmartPeg type selected for the Implant See SmartPeg ref...

Страница 11: ...t the local sales representative or distributor for further instructions Difficult to measure in an exact recommended directionn No space e g due to adjacent teeth Try to measure at a slightly different angle Difficulties attaching the SmartPeg Wrong SmartPeg Ensure that the SmartPeg is compatible with the implant system 18 Service and Support In the event of an instrument malfunction contact the ...

Страница 12: ...Osstell AB Stampgatan 14 411 01 Göteborg Sweden Phone 46 31 340 82 50 info osstell com www osstell com Osstell AB Date of issue 2018 05 25 25087 00 ML OSSTELL Beacon USER MANUAL ...

Отзывы: