  Do not expose the instrument to extreme high temp, e.g. leaving 

it in the car dashboard on a warm sunny day.

  The Osstell Beacon is not protected from ingress of fluids, e.g. 

water, at the USB connector (IP20 classified).

  Mains-operated power supplies or USB cable used for charging, 

shall not be reachable by the patient.

Precautions: 

  To avoid interference with other equipment, the instrument 

should not be held close to electronic devices.

  Only use the acceptable cleaning fluids listed in section 15 

 

when cleaning the instrument.

  Do not use the instrument in the presence of explosive or  

flammable materials.

  See section 4, 5 and 10 for information about approved and 

compatible accessories.

3) Intended Use

Osstell Beacon is intended for use as a Dental Implant Stability

Analyzer.

4) Indication for use

Osstell Beacon is indicated for use in measuring the stability of

implants in the oral cavity and craniofacial region.

Conditions

Surgically placed implants or abutments for which there is space  

to attach a compatible SmartPeg.

Reasons for use

The Osstell Beacon can add important information to the evaluation 

of implant stability and can be used as part of an overall treatment 

evaluation program. The final implant treatment decisions are the 

responsibility of the clinician.  

Contraindications

 

Osstell Beacon is contraindicated for implant systems to which the 

SmartPeg could not be attached for mechanical incompatibility 

reasons. See section 10 for more information about SmartPegs.

Osstell Beacon is contraindicated when used together with Pegs not 

approved by Osstell. Osstell Beacon is contraindicated where it is 

not possible to attach the SmartPeg due to lack of space, or where it 

impinges on other artificial or anatomical structures.

5) Description

The Osstell Beacon is a handheld instrument that involves the use  

of the non-invasive technique, Resonance Frequency Analysis.  

The system involves the use of a SmartPeg attached to the dental 

implant or abutment by means of an integrated screw. The Smart-

Peg is excited by a magnetic pulse from the instrument tip.
The resonance frequency, which is the measure of implant stability, 

is calculated from the response signal. Results are displayed on the 

instrument as the Implant Stability Quotient (ISQ). The ISQ is scaled 

from 1 to 100. It is a measurement of the stability of the implant 

and is derived from the resonance frequency value obtained from 

the SmartPeg. The higher the number, the greater the stability. The 

instrument software can be updated by using the Osstell USB cable. 

Содержание Beacon

Страница 1: ...osstell com USER MANUAL Osstell Beacon GB ...

Страница 2: ...rm a measurement in two directions Buccal lingual and Mesial Distal as guided by the instrument This is impor tant to detect the lowest implant stability The SmartPegs are disposable and should only be used for one or multiple measurements at one treatment session for a single patient only to avoid cross contamination Repeated re use may result in false readings due to wear and tear of the soft al...

Страница 3: ...verall treatment evaluation program The final implant treatment decisions are the responsibility of the clinician Contraindications Osstell Beacon is contraindicated for implant systems to which the SmartPeg could not be attached for mechanical incompatibility reasons See section 10 for more information about SmartPegs Osstell Beacon is contraindicated when used together with Pegs not approved by ...

Страница 4: ...lectronic product is approved by Federal Communications Comission FCC FCC ID QOQBLE113 IC 5123A BGTBLE113 Non ionizing electromagnetic radiation Not Sterilizable Temperature limitation Humidity limitation Atmospheric pressure limitation 6 Safety Symbols 1 2 3 4 5 6 Your Osstell Beacon system includes the following items 1 Osstell Beacon Instrument 2 Osstell USB cable 3 Osstell SmartPeg Mount 4 Oss...

Страница 5: ...acon Instrument The Osstell Beacon is activated by a built in motion sensor As soon as the motion sensor detects movement the Osstell Beacon will start up flash red yellow green and shortly show battery status in the lower display and then be ready for measurement in the BL Buccal Lingual direction which is indicated in the upper display The Osstell Beacon will start to measure indicated by an aud...

Страница 6: ...implant placement to get a baseline for future measurements throughout the healing process Before the final restoration another measurement is taken which makes it possible to observe the stability development of the implant It is recommended to measure in both Buccal Lingual and Mesial Distal direction to find the lowest stability Therefore the Osstell Beacon prompts the user to measure in both t...

Страница 7: ...ould be taken on the implant before the abutment is attached and then a second measurement on the abutment 13 Interpret the Result Implant Stability An implant can have different stability in different directions The stability of the implant is dependent of the surrounding bone configuration There is often a direction where the stability is lowest and a direction where the stability is highest and...

Страница 8: ...ze or soft cloth with an acceptable see list below surface disinfectant and wipe the whole Osstell Beacon instrument Note Do not autoclave the Osstell Beacon instrument Routinely check the surface of the instrument tip and overall surface for possible cracks and residuals Acceptable Cleaners Isopropyl alcohol Ethyl alcohol based cleaners Do not use Strongalkalidetergentofanytype includinghandsoaps...

Страница 9: ...mbols used follow the European standard EN 60601 1 and ISO standards 9687 and 15223 as far as possible Notes on electromagnetic compatibility EMC Medical electrical equipment is subjected to particular precautions with regards to EMC and must be put into operation in accordance with the EMC notes included below Osstell guarantees the compliance of the device with the EMC requirements only when use...

Страница 10: ... to 1060 hPa IP class IP20 17 Troubleshooting No measurement or unexpected value Re Used SmartPeg The SmartPegs are disposable and should only be used for one or multiple measurement at one treatment session for a single patient only Repeated re use may result in false readings due to wear and tear of the soft aluminium SmartPeg threads Wrong SmartPeg type selected for the Implant See SmartPeg ref...

Страница 11: ...t the local sales representative or distributor for further instructions Difficult to measure in an exact recommended directionn No space e g due to adjacent teeth Try to measure at a slightly different angle Difficulties attaching the SmartPeg Wrong SmartPeg Ensure that the SmartPeg is compatible with the implant system 18 Service and Support In the event of an instrument malfunction contact the ...

Страница 12: ...Osstell AB Stampgatan 14 411 01 Göteborg Sweden Phone 46 31 340 82 50 info osstell com www osstell com Osstell AB Date of issue 2018 05 25 25087 00 ML OSSTELL Beacon USER MANUAL ...

Отзывы: