Potentiostat
User Manual
Ossila.com
13
Ossila Limited © 2021
9.3.1
Voltammetry Settings
Figure 9.2.
Linear Sweep and Cyclic Voltammetry measurement settings.
(I)
Connected Systems
•
Select the COM port of the connected unit you intend to use.
I.
This box will be populated automatically with the addresses of any units connected
to the computer when the software starts.
II.
To rescan for connected units (in case the connection is changed) click the refresh
icon next to the drop-down box.
(II)
Current Range
•
Select the range of currents to be used for the measurement or automatic range selection.
I.
This defines the upper limit, accuracy, and resolution of the current measurements
that can be performed by the system.
II.
Automatic range selection will start on the lowest current range and automatically
switch to higher ranges if the current increases above the maximum for a range.
(III)
Voltammetry Type
•
Select whether to perform Linear Sweep or Cyclic voltammetry.
(IV)
Scan Rate (mV/s)
•
The rate at which the potential will be changed during the scan, measured in millivolts per
second.
(V)
Start Potential (V)
•
The potential in volts at which the measurement starts.