22
9.
Maintenance
9.1
Cleaning
Maintenance consists of periodic cleaning. The exterior of the instrument can be cleaned with
a clean, dry cloth to remove any oil, grease, or grime. Never use liquid solvents or detergents.
Repairs or servicing not covered in this manual should only be performed by qualified
personnel.
9.2
Repair and Service
There are no user-serviceable parts in this unit except for the fuse which is accessible
externally. If the unit is faulty, return it to Ossila Limited. Our service department will promptly
quote to repair any faults that occur outside the warranty period.
9.3
Storage Conditions
The UV Ozone Cleaner should be kept in dry conditions; away from direct sources of heat or
sunlight, and in such a manner as to preserve the working life of the instrument.
10.
Troubleshooting
Problem
Possible cause
Action
No power /
display
a.)
The power switch on the
unit is in the OFF position
b.)
The power supply may
not be connected
properly
c.)
The fuse on the rear
panel has blown
d.)
If the unit was supplied
with a transformer, it
may be faulty
e.)
Fault on circuit board
a.)
Check the connection and
ensure the power is turned
ON
b.)
Ensure the unit is firmly
plugged in to the power
supply, and the plug is
firmly connected to both
the adapter and the
working power socket
c.)
Ensure the unit is
unplugged. Check the fuse
on the rear panel. If it has