USER MANUAL FOR THE OSMIO INFINITY HYDROXY GAS/WATER PRODUCTION ELECTROLYSIS SYSTEM
31
6.2 Cleaning Nasal Cannula and Tubing (home use)
6.2.1 Cleaning cannula and tubing
Nasal cannulas and tubing require to be cleaned after each use (or daily if system used
several times a day).
To sanitise using Osmio Sanser 300 mg/l solution
1.
Prepare 0.5 litre of solution (10 min cycle) see manual for instructions.
2.
Leave your cannula in this solution for 5 minutes.
3.
Rinse the nasal cannula thoroughly, shake off water and hang it up to dry.
4.
Allow the cannula to dry completely before use.
Hypertonic solution can be used to clean cannula and tubing. Dissolve 1 teaspoon of
salt and ½ teaspoon of baking soda in 1L of cooled boiled water. Allow 30 minutes
contact time.
If you are using any other sanitising solution, please refer to manufacturer for contact
time and safety instructions.
If you have been unwell, please replace the cannula with a new one. Each user should
have their own cannula.
6.2.2 Cleaning Humidifier and bubbler
To sanitise using Osmio Sanser 300 mg/l solution
1.
Prepare 0.5 litre of solution (10 min cycle) see manual for instructions.
2.
Rinse bubbler and humidifier with soapy water
3.
Divide solution between bubbler and humidifier, screw on lids and shake well
4.
Leave for 5 minutes, rinse thoroughly,
5.
Refill with 0TDS water
Hypertonic solution can be used to clean bubbler and humidifier. Dissolve 1 teaspoon of
salt and ½ teaspoon of baking soda in 1L of cooled boiled water. Allow 30 minutes
contact time. Remove bubbling stone from the bubbler if using this method.