background image

Series3 Auto User Guide

15

Problems

Reasons

Maintenance

Mattress fail 

to inflate or 

do not inflate 

completely.

Pump 

issue

1.  Pump does not work.

1.  After powered on, check if visible LED 

light turns on. If not, please check the 

below issues: 

1.1  Check if power cord is plugged into 

appropriate voltage AC outlet. 

1.2  Check if fuse is loose or burned out, 

make sure it’s connected well or 

replace with new one.  

1.3  Open the pump and see if wires inside 

are connected well, make sure they are 

not loose.

1.4  Change lower PCB.  

2.  Air pressure from 

pump is too low.

1.  Check if air pressure and air flow 

(100mmHg, 8.0L) are high enough 

from the compressor, if not then 

replace with new compressor. 

2.  Check if there is air leakage from the 

exchanger, if yes then replace with new 

exchanger. 

3.  Check if silicone tube inside the pump 

is connected well.

  

Mattress 

issue

1

.  Quick connector on 

mattress does not 

connect well with 

pump. 

2.  Air tube connected 

to T/L connector 

and air valve is loose, 

CPR connector is not 

capped. 

3.  One way valve is 

broken. 

4.  Air cell is leaking.

 

1.  Make sure quick connector on 

mattress is connected well with pump. 

2.  Make sure T/L connector and air valve 

is connected well, CPR connector is 

capped well. 

3.  Change one way valve. 
4.  Change air cell.

Troubleshooting guide

Problems

Reasons

Maintenance

Mattress has pillow 

function, but air cell fail to 

inflate.

 

1.  One way valve is assembled 

reversely.

1.  Assemble the one way valve 

in correct direction.

Pump is working but 

synchronous motor does 

not work; thus mattress 

does not alternate, and 

alternate failure alarm is 

activated.

1.  Synchronous motor is out of 

order. 

2.  Wires inside synchronous motor 

not connect well. 

3.  Lower PCB is out of order. 

1.  Change synchronous motor
2.  Make sure wires are 

connected well. 

3.  Change the lower PCB.

Pump and motor keep 

working, but cycle time is 

incorrect. The alternate 

failure alarm is activated.

1.  Micro switch on the exchanger 

is out of order.

2.  Lower PCB is out of order.

1.  Change the micro switch.
2.  Change the lower PCB.

When powered on, 

compressor stop after 

working some time; but 

the exchanger keep rotary.

1.     Pressure detector is out of 

order.

1.  Change the lower PCB. 

Mattress pressure is low 

but alarm is not activated. 

1.  Pressure detector is out of 

order.

1.  Change the lower PCB. 

Push button on panel is 

not operated well, and 

LED indicator does not 

light up.

 

1.  Push button is not operated 

well.  

2.  LED is out of order.  

1.  Change the upper PCB. 

Mattress pressure is too 

high or too low.

1.  Pressure sensor is out of order.

1.  Change the lower PCB.  

Power failure alarm can’t 

be activated after power 

failure. 

1.  Battery is out of order.

1.  Change the lower PCB. 

 

Troubleshooting guide continued

03

 

Know your symbols

04

 

Important safeguards

04

 Statements

05

 Introduction

05

 

Package contents

06

 

Product function

08

 

System installation

09

 Operation

11

 

Cpr function

11

 

Pressure set up

11

 

Low pressure warning

12

 Cleaning

13

 

Environment requirements

13

 

Handling and storage

13

  

Maintenance

14

 Troubleshooting

16

 

Specifications

Содержание 3 Auto Series

Страница 1: ...Version 3 17 11 2022 User Guide Series3 Auto ...

Страница 2: ...ure warning 12 Cleaning 13 Environment requirements 13 Handling and storage 13 Maintenance 15 16 Troubleshooting 17 Specifications Know your symbols Manufacturer Do not iron Foot end Tumble dry CE marked Do not dry clean REF Catalogue number Resistant to ignition REP EC Authorized EC Representative Wipe down Machine wash at 95 Recycling Drip dry UKCA marked 1 CL 1 Chlorine Medical Device Airflo Xi...

Страница 3: ...a properly grounded outlet only Do not disassemble repair or rebuild the product without permission otherwise may has risk of fire and danger in life Grounding Before any connection is made make certain that this product shall be connected to a protective earth conductor via the three wire power cord the three blade plug shall be inserted only into a socket outlet that provides a protective earth ...

Страница 4: ...on for 3 seconds to lock the panel press again for 3 seconds to unlock the panel Power Switch 1 The switch is at the right side of the pump Turn ON OFF the power and press Power button on the panel the pump will start stop operation Press Cycle Time 2 The cycle time can be selected from the panel The cycle time value options are 10 15 20 or 25 minutes Auto weight 3 Auto Press the button the pump w...

Страница 5: ... under each end of your bed Ensure buckles are securely fastened and straps are pulled tight c IMPORTANT On a profiling bed secure the side straps around the moveable sections of the bed base DO NOT secure to the side rails or main bed frame irreparable damage could be caused to the bed and mattress d Confirm there are no sharp objects in the immediate area which may risk damage to the Mattress Re...

Страница 6: ...le pressure please refer to page 8 for the hand check procedure CPR function When there is an emergency requirement to perform CPR on the patient press the CPR button at the head section of the mattress and disconnect the air hose from the pump to release the air quickly from the mattress Pressure set up The pump unit can automatically set the pressure according to patient s weight Other functions...

Страница 7: ...e to disconnect all the air cells one by one and spray the disinfectant on all sides including all the connecting tubes and hoses Let it sit for least 10 minutes If there is a base after you remove all the air cells the base has to be sprayed down with the disinfectant inside and outside Let it sit for the required incubation time and wipe down with a cloth Repeat the process with the tubing set s...

Страница 8: ...Change air cell Troubleshooting guide Problems Reasons Maintenance Mattress has pillow function but air cell fail to inflate 1 One way valve is assembled reversely 1 Assemble the one way valve in correct direction Pump is working but synchronous motor does not work thus mattress does not alternate and alternate failure alarm is activated 1 Synchronous motor is out of order 2 Wires inside synchrono...

Страница 9: ...able Nylon PVC 840D Case Material Flame Retardant ABS 20 x 8 Cell on Cell air cells Max firm mode CPR valve for emergency procedures Alarm Low pressure Power failure Alternate failure Pillow function Alternate Static Seat inflate Auto weight mode Quick connectors Size 31cm L x 12 5cm W x 21cm H Maximum Weight Capacity 200 kg Weight 2 5 kg Note The above specifications are also applicable to those ...

Страница 10: ...Series3 Auto User Guide 19 Notes Notes ...

Страница 11: ...Series3 Auto User Guide Edward House 5 Penner Road Havant PO9 1QZ 02394 318 318 ask oska uk com oska uk com Pressure Care Experts ...

Отзывы: